Hach-Lange HQD - Gel filled ORP_Redox electrode MTC User Manual

User instructions, Safety information, Overview

Advertising
background image

1

DOC272.53.80033

User instructions

Gel filled ORP/Redox electrode: Model MTC10101, MTC10103,
MTC10105, MTC10110, MTC10115 or MTC10130

Safety Information

Precautionary Labels
Read all labels and tags attached to the instrument. Personal injury or damage to the
instrument could occur if not observed. A symbol, if noted on the instrument, will be included
with a danger or caution statement in the manual.

Overview

The MTC101 series is a non-refillable, gel-filled combination oxidation reduction potential
(ORP/Redox) electrode with a built in temperature sensor. The MTC10101 or MTC10103
electrode is available with a 1 or 3 meter cable and is intended for laboratory use. The
MTC10105, MTC10110, MTC10115 or MTC10130 electrode is available with a 5, 10, 15 or 30
meter cable and is intended for field use. The electrode gives direct measurements of absolute
mV values in wastewater, drinking water and general applications.

Specifications

Specifications are subject to change without notice.

Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after
12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC),
European electrical equipment users must now return old or end-of life equipment to the Producer for disposal at no
charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return
end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.

This is the safety alert symbol. Obey all safety messages that follow this symbol to avoid potential injury. If on the
instrument, refer to the instruction manual for operation or safety information.

Specification

Details

Electrode type

Digital combination electrode with a non-refillable Ag/AgCl reference and a built-in
temperature sensor

Range

± 1200 mV

Resolution

0.1 mV

Temperature accuracy

± 0.3 ºC (± 0.54 ºF)

Operating temperature range

0 to 80 °C

Storage temperature range

5 to 40 °C

Junction

Open junction

Reference potential versus
Standard Hydrogen Electrode

207 mV at 25 ºC

Reference type

Ag/AgCl (3M KCl)

Minimum sample depth

20 mm

Dimensions

Standard: Diameter: 12 mm (0.47 in.), Length: 103 mm(4.056 in.), Total length: 200 mm
(7.9 in.), Cable length: 1 or 3 meter (3.3 or 9.8 ft). Rugged: Diameter 12 mm (0.47 in.),
Length: 20 mm (0.78 in.), Total length: 220 mm (8.7 in.), Cable length: 5, 10, 15 or 30
meter (16.4, 32.8, 49.2, or 98.4 ft).

Cable connection

Digital output and connector compatible with HQd meters

Warranty

Sensor is covered by a 6 month warranty.