Figure 1 Probe overview

1 Probe soaker bottle

4 Protective tape and filing-hole

2 Reference junction, glass bulb and temperature sensor

5 1 or 3 meter (3.28 or 9.84 ft) cable

3 Filling-hole cap

6 Probe soaker bottle holder

Preparation for use

Prepare the probe for use before calibration or sample measurement.

1. Turn the probe soaker bottle cap counter-clockwise to loosen the cap.
2. Remove the soaker bottle from the probe.
3. Rinse the reference junction and glass bulb with Ionic Strength Adjustor (ISA) rinse

solution (refer to

Prepare the ISA rinse solution

on page 11). Blot dry with a lint-free

cloth.

4. Remove the protective tape from the filling hole before initial use (refer to

Figure 1

on page 2). Dispose of the protective tape.

5. Add filling solution to the probe as necessary (refer to

Fill the probe

on page 11).

The filling solution must be above the standard solution or sample level during
measurement (

Figure 2

).

6. Make sure that the filling hole is open during measurement for the proper flow of the

filling solution.

Note: After long-term storage, the probe must be conditioned in 25 mL of 100 mg/L Na

+

standard solution with the contents of one Sodium ISA powder pillow for at least 30 minutes

prior to use. If probe stabilization is slow after storage, condition the probe for up to one hour in

25 mL of 100 mg/L Na+ standard solution with the contents of one Sodium ISA pillow.