Message or symptom

Possible cause

Action

Slow stabilization time

Tape not removed from the
filling-hole

Remove the tape.

Contamination in the glass
sensor

Clean the probe. Refer to

Clean the probe

on page 10.

Contamination in the filling
solution

Drain and replace the filling solution with fresh solution.

Filling hole closed

Make sure that the filling hole is open and above the
solution level.

Poor contact between reference
junction and solution

Shake the probe in the solution from side to side to
refresh the reference junction.

LIS sample

Soak the probe in the sample for 10 to 15 minutes before
the initial measurement or use a LIS chamber or flow-
through cell.

Low sample temperature or
temperature difference between
samples

Examine the sample temperature. The lower the
temperature or the higher difference of temperatures
between samples, the longer the stabilization time will
be.

Air bubbles around inner
reference electrode

Carefully tap the probe with hand or shake the probe
downward to remove any air bubbles.

Dehydrated glass bulb

Rehydrate the probe. Refer to

Storage

on page 11.

Not sufficient electrolyte flow

1

Open the sleeve junction.

pH probe too old (membrane
degradation)

Replace the probe.

Calibration errors

Calibration not done correctly

Calibrate again with fresh pH buffers.

Contaminated glass sensor

Clean the probe. Refer to

Clean the probe

on page 10.

Slope exceeds the criteria for %
of theoretical (as defined in the
method slope limit)

1.

Adjust the slope limits by changing the PHC735
calibration settings and method.

2.

Calibrate again the probe.

3.

Make a check standard to examine the performance
of the probe.

13