Troubleshooting, If a measurement error occurs, refer to – Hach-Lange ISENH318103 User Manual User Manual

Page 11

Advertising
background image

Note: After long-term storage, the ISENH3181 probe (with membrane module assembled)

might need to be conditioned in Ammonia probe storage solution for up to 30 minutes to

improve the stabilization speed.

Troubleshooting

Message or symptom

Possible cause

Action

Probe not supported

Software not updated

To download the most current version of the
software, refer to the applicable product page on the
manufacturer's website.
Refer to the HQd Series meter manual for specific
instructions for the meter model.

HQd meter does not support
IntelliCAL

®

probe

Contact a Technical Support Representative.

Connect a probe or probe
requires service

Probe not connected properly

Disconnect, then connect the probe. Tighten the
locking nut.

Software not updated

To download the most current version of the
software, refer to the applicable product page on the
manufacturer's website.
Refer to the HQd Series meter manual.

Large number of methods stored on
probe.

Continue to let probe connect. Do not disconnect
probe.

Damaged probe

Make sure connectivity with another probe or meter
to confirm isolated issue with probe. Contact a
Technical Support Representative.

mV reading is the same for all
solutions

Soaker bottle not removed

Remove the soaker bottle.

Electrical issue

Contact a Technical Support Representative.

Slow response time

Dirty sensing element

Clean the probe (body, membrane module and glass
bulb). Refer to

Clean the probe

on page 10.

Membrane failure

Replace the membrane module.

Dirty filling solution

Replace the filling solution.

Low sample temperature or
temperature difference between
samples

Check the sample temperature. The lower the
temperature or the greater the difference of
temperatures between samples, traditionally the
longer the response time.

Bubbles trapped under sensor tip

Gently shake the probe until bubbles are removed
from under sensor tip.

Slope out of range (refer to

Check probe response

on page 13)

pH is incorrect

Make sure the pH is > 11 after each ISA addition.

Ionic strength adjustor (ISA) not
used

Add ISA to each sample and standard (one powder
pillow per 25 mL of solution).

Insufficient conditioning

Condition for at least in ammonia probe storage
solution.

Damaged probe

Contact a Technical Support Representative.

Incorrect standards

Calibrate using freshly prepared standards.

Dirty sensing element

Clean the probe (body, membrane module and glass
bulb) and recalibrate.

Bubbles trapped under sensor tip

Gently shake the probe until bubbles are removed
from under sensor tip.

11