Hach-Lange PHC20103 User Manual User Manual

User manual, Safety information, Specifications

Advertising
background image

User Manual

DOC022.52.80197

General use pH probe: Models PHC20101, PHC20103

Safety information

Precautionary labels

Read all labels and tags attached to the instrument. Personal injury or damage to the
instrument could occur if not observed.

Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after 12
August of 2005. In conformity with European local and national regulations (EU Directive 2002/98/EC), European
electrical equipment users must now return old or end-of-life equipment to the Producer for disposal at no charge to
the user.

Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life
equipment, producer-supplied electrical accessories, and all auxillary items for proper disposal.

Specifications

Note: Specifications are subject to change without notice.

Specifications

Details

Probe type

Digital combination non-refillable, semi-liquid gel-filled probe with double
junction reference and built-in temperature sensor

pH range

pH 0 to 14

pH resolution

User-selectable stabilization time and resolution—Fast: 0.1, Fast: 0.01,
Medium: 0.01, Slow: 0.01 or Slow: 0.001

Slope

-59 mV/pH (85 to 115% at 25 °C (77 °F) per Nernstian theoretical value)

Operating temperature range

0 to 80 °C (32 to 176 °F)

Storage temperature range

5 to 40 °C (41 to 104 °F)

Junction

Ceramic

Reference type

Ag/AgCl

Sodium (alkalinity) error

<20% at pH 14 in 1 M NaCl

Temperature accuracy

±0.3 °C (±0.54 °F)

Minimum sample depth

15 mm (0.59 in.)

Dimensions (standard)

Diameter: 12 mm (0.47 in.)
Length: 103 mm (4.056 in.)
Total length: 200 mm (7.9 in.)
Cable length: 1 or 3 m (3.28 or 9.84 ft)

Cable connection

M12 digital output and connector compatible with HQd meters

Product overview

The PHC201 series probe is a non-refillable, semi-liquid gel-filled combination pH probe
with a built-in temperature sensor (

Figure 1

). The standard PHC20101 or PHC20103 probe

comes with a 1 or 3 m (3.28 or 9.84 ft) cable and is intended for laboratory use. The probe
provides pH measurement in general aqueous applications. The probe is not suitable for
use with organic solvents, low ionic strength, wastewater, high solids or high alkaline sample
types.