Hach-Lange 5341 User Manual

Page 5

Advertising
background image

Measuring range

0 to 14 pH

Operating temp.

0 to 80 °C

Max. pressure

6 bar

Reference element

cartridge

Diaphragm

ceramic

Electrolyte

gel

Body material

glass

Thread

PG 13,5

Connector

S8

Min. immersion depth

17 mm

5

GB

... impossible to perform calibration.
... very slow response.

Dirt on the diaphragm or the mem-
brane.
De-hydrated membrane.
Electrode ageing or wear.

Clean, regenerate, replace the elec-
trode.

Storage
The connector should be protected from
humidity by the protective cap

.

Place the electrode in the storage protec-
tor

with KCl 3M.

Specifications

Applications
Water containing fluoride.

Warning. The resistance of the

membrane is directly related to the HF
concentration and the pH of the sample
(see ).

Limitations
Viscous solutions or samples with colloids.
Distilled water.

CERTIFICATE OF QUALITY

The attached electrode 53 41 has passed
the quality test and follows the next speci-
fications:
• Asymmetry potential < ± 15 mV.
• Sensitivity, pH 4 to 7 (at 25 ˚C) > 98%.
• Response time, pH 4 to 7 < 20 s.

53 41_DOC013.97.90280_Ed 0510.indd 5

3 41 DOC013 97 90280 Ed 0510 i dd 5

23/9/10 10:29:58

23/9/10 10 29 58