Advanced operation, Change measurement options – Hach-Lange ISENO318103 User Manual User Manual

Page 7

Advertising
background image

value of the check standard and either Check Standard Passed or Check Standard
Failed.

5. If the display shows Check Standard Passed, the check standard measurement is

within the accepted limits set by the administrative user. Select Done to continue with
the sample measurement.

6. If the display shows Check Standard Failed, the measurement is outside of

accepted limits set by the administrative user and a recalibration is recommended. If
the acceptance criteria is set to Cal Expires on Failure: Yes, the display shows the
calibration icon and a question mark until the probe is recalibrated. To correct the
probe calibration and status indicator, calibrate the probe (refer to

Calibration

on page 2).

Advanced operation

Parameter-specific settings can be changed through the Full Access Options menu.
Details about menu navigation, available options and how to change them are given in
the screens, tables and procedures throughout this section.

The settings that can be changed are shown in

Table 2

.

Table 2 Parameter-specific settings

Setting

Options

Measurement Options

Units

Significant digits

Auto stabilization

Stability criteria

Upper and lower range limits

Calibration Options

Standard set

Calibration units

Minimum calibration points

Slope limit

Calibration reminder

Check Standard Options

Standard

Check standard reminder

Acceptance criteria

Change measurement options

Methods are groups of factory-set or user-defined settings relevant to specific
applications. If the meter is set to a factory-set method and the Modify Current Settings
option is chosen, a prompt for a new name is shown after the changes are entered. The
settings are saved with this name to distinguish them from the factory-set methods, which
cannot be changed. A saved method can be used instead of multiple adjustments to the
individual settings. Changes made to a user-defined method are automatically saved with
the existing name. Multiple methods can be saved for the same probe on each meter.

7