8. Error messages

Display

Caused by

Action

- Measurement of D.O. or temperature out of range.
- In calibration:

- Standard 0%: limits <0 nA and >2nA

- Standard 100%: limits <30 nA and >110 nA

Check standards.
Check sensor.

Low battery charge.

Replace the batteries.

E1

Unstable reading during a stability measurement or
during calibration.
180-second timeout.

Check sensor.
Check polarisation.
Check flow.

E2

Exceeded upper limit of sensor current during

measurement

250 nA

Check sensor.