Specification

Details

pH

Calibration

Up to five points: 1.68, 4.01, 6.86, 7.00, 9.18, 10.01, 12.45

Accuracy

±0.01 pH

Resolution

0.01 pH

Range

-2.00 to 19.99

mV

Calibration

None

Accuracy

±1 mV

Resolution

Autoranging, 0.1 and 1

Range

Autoranging, ±199.9 mV to ±1999 mV

Temperature

Calibration

None

Accuracy

±0.5 °C

Resolution

0.1 °C (0.1 °F)

Range

-5 to 105 °C (23 to 221 °F)

ISE

Calibration

Up to five points

Accuracy

Probe dependent

Resolution

0.1 ppm–0.1 ppt

Range

Autoranging, -0.0 ppm to 1999 ppt

Conductivity

Calibration

Up to five points

Accuracy

±1% full scale or ±1 digit

Resolution

0.01 µS, 0.1 µS, 1 µS, 0.01 mS, 0.1 mS

Range

Autoranging: 0.00 to 19.99 µS, 20.0 to 199.9 µS, 200 to 1999 µS, 2.00 to 19.99 mS,
20.0 to 199.9 mS

TDS

Calibration

Up to five points

Accuracy

±1% full scale or ±1 digit

Resolution

0.01 ppm, 0.1 ppm, 1 ppm, 0.01 ppt, 0.1 ppt, 1 ptt, 0.1 mg/L, 1 mg/L, 0.01 gal/L,
0.1 gal/L

Range

Autoranging, ppm: 0.00 to 9.99 ppm,10.0 to 99.9 pm,100 to 999 ppm,1.00 to 9.99 ppt,
10.0 to 99.9 ppt,100 to 200 ppt
mg/L: 0.00 to 199.9 mg/L, 200 to 1999 mg/L, 2.00 to 19.99 gal/L, 20 to 50 gal/L

Salinity

Calibration

None (derived from conductivity)

Accuracy

±0.1 ppt (–2 to +35 °C or 28.4 to 95 °F)

Resolution

0.1 ppt, 1%