Hach-Lange SENSION+ MM374 User Manual

Page 3

Advertising
background image

Table of contents

Specifications

on page 3

Advanced operation

on page 12

General information

on page 4

Data logger

on page 13

Installation

on page 5

Maintenance

on page 15

User interface and navigation

on page 8

Troubleshooting

on page 17

Startup

on page 9

Replacement parts and accessories

on page 18

Standard operation

on page 9

Standard solutions

on page 19

Specifications

Specifications are subject to change without notice.

Specification

Details

Dimensions

35 x 20 x 11 cm (13.78 x 7.87 x 4.33 in.)

Weight

1100 g (2.43 lb)

Meter enclosure

IP42

Power requirements
(external)

100–240 V, 0.4 A, 47-63 Hz

Meter protection class

Class II

Pollution degree

2

Installation category

Category II

Altitude requirements

Standard 2000 m (6562 ft) ASL (Above Sea Level)

Storage temperature

–15 to +65 °C (5 to +149 °F)

Operating temperature

0 to 40 °C (41 to 104 °F)

Operating humidity

< 80% (non-condensing)

Specification

Details

Resolution

pH: 0.1/0.01/0.001, ORP: 0.1/1 mV, ISE:
programmable, temperature: 0.1 ºC (0.18 ºF), EC:
variable, resistivity: variable, NaCl: variable, TDS:
variable

Measuring error (±
1 digit)

pH: ≤ 0.005, ORP: ≤ 0.2 mV, temperature: ≤ 0.2 ºC (≤
0.36 ºF), EC: ≤ 0.5%, resistivity: ≤ 0.5 %, NaCl: ≤
0.5 %, TDS: ≤ 0.5%

Reproducibility (±
1 digit)

pH: ± 0.001, ORP: ± 0.1 mV, temperature: ± 0.1 ºC (±
0.18 ºF), EC: ± 0.1%, resistivity: ± 0.1%, NaCl ± 0.1%,
TDS ± 0.1%

Data storage

330 results and last 9 calibrations

Connections

2 Combined or indicator probe: BNC connector (Imp.
>10

12

Ω); 2 Reference electrodes: banana connector;

A.T.C. type Pt 1000: banana or telephonic; 2 magnetic
stirrers: RCA connector
Conductivity probe with built-in Pt1000 sensor (or NTC
10 kΩ probe): telephonic connector
RS232C for printer or PC: telephonic connector;
external PC keyboard: mini DIN connector

Temperature correction Channel 1 pH: Pt 1000 temperature probe (A.T.C.),

NTC 10 kΩ probe, manual, Channel 2 temperature,
isopotential pH programmable, standard value 7.00,
CE: Pt 1000 temperature probe (A.T.C.), linear
function, TC=0.00 to 9.99%/Temperature, TRef: 20°C
(68 °F) or 25°C (77 °F), non-linear function for natural
waters (UNE EN 2788) Channel 2 pH: Pt
1000 temperature probe (A.T.C.), manual, NTC 10KΩ
probe, isopotential pH programmable, standard value
7.00

Measurement display
lock

Continuous measurement, by stability, by time

Display

Liquid crystal, backlit, 128 x 64 dots

Keyboard

PET with protective treatment

Certification

CE

English 3