3.7 Setting the temperature of the measuring shaft

Before the instrument leaves the factory the temperature control of the measuring shaft

is set to 15

°

C. The temperature control of the measuring shaft can be switched on or

off.

The temperature control can be adjusted. To do this, measure the measuring shaft

temperature with an accurate thermometer. (Introduce about 1 ml water into the

measuring cuvette and place the cuvette in the measuring shaft. Measure the water

temperature after the temperature adjustment.) In the adjustment program, enter the

measured temperature via <F1> or <F2> as the actual temperature. The program

calibrates the temperature control so as to reach the required temperature of 15

°

C.

3.8 Setting the screen contrast

The screen contrast can be individually adjusted.

Sequence:

Select the option <temp> or <cont>.

Left:

Switch the temperature control on

or off with F2.

Adjust the measuring shaft

temperature with F1.

right:

Set the screen contrast.

options

temp cont srv

temp: 15.1°C on
just off

Kontrast

-

+

meas. :15.0°C

+

-