Installation

Fig. 3-20

Terminal assignments

Table 3-1 Terminal assignments

1.

Mains cable (power supply)

3.

Heater connecting cable, sample tube

2.

Earth connection

4.

Heater connecting cable, suction tube

Sicherungen/Fuses

250 V

F1:
F2:
F3:
F4:

T3,15A L
T3,15A L

T6, 3A L
T6, 3A L

T4A E
T4A E
T7A E
T7A E

IEC

UL

N O T I C E

Route the signal lines (21, 22, 23, 24) separated from the power cable.

Terminal no.

Connection

1

N (230 V AC / 50-60 Hz, option: 115 V AC)

2

L (230 V AC / 50-60 Hz, option: 115 V AC)

7

N (sample tube heater)

8

N (suction tube heater)

9

L (sample tube heater)

10

L (suction tube heater)

18

Earth for screening signal wires

19

Earth for suction tube heater

20

Earth for sample tube heater

21/22

Floating fault signalling contacts (normally open)

23/24

Floating warning contacts (normally open)