Standard Operations

5.4.4.1 Take single wavelength measurements (single reading)

1. Insert the blank cuvette into the cuvette/sample cell holder.

Press Zero.

Note:The Read key is only active after the zero measurement has

been carried out.

2. Insert the sample cuvette into the cuvette/sample cell holder.

Press Read.

Note: For data storage, see

section 5.3.2 on page 38

.

5.4.4.2 Take single wavelength measurements (continuous readings)

1. Insert the blank cuvette into the cuvette/sample cell holder.

Press Zero.

Note: In the reading mode ”Continuous“ only the Zero key is shown to
start the reading. The reading sequence is started automatically.

2. Insert the sample cuvette into the cuvette/sample cell holder.

3. Press Options and then the Store icon to store the displayed

data in the Data Log.

Note: For data storage, see

section 5.3.2 on page 38

.