2 data table – Hach-Lange PCII-Ammonium LCK 303_304_305 User Manual

Page 10

Advertising
background image

System start up

10

This application instruction contains the calibration data of the HACH-LANGE
Cuvette Tests for Ammonium-Nitrogen/Ammonium:

LCK 303: 2-47 mg/l NH

4

-N or 2.5-60 mg/l NH

4

LCK 304: 0.015-2 mg/l NH

4

-N or 0.02-2.5 mg/l NH

4

LCK 305: 1-12 mg/l NH

4

-N or 1.2-15 mg/l NH

4

Select the calibration data of maximum two measuring ranges which will be the
suitable one.

Each programming procedure has only to be followed once prior to the first test
evaluation.

For detailed information please read the POCKET COLORIMETER II instruction
manual
.

3.2.2

Data table

3.2.2.1

Evaluation as Ammonium-Nitrogen NH

4

-N

3.2.2.2

Evaluation as Ammonium NH

4

Table 1 Evaluation as Ammonium-Nitrogen NH

4

-N

LCK 303

LCK 304

LCK 305

RES (resolution)

--.--

-.---

--.--

S0 (standard 0)

02.00

0.015

00.99

A0 (absorbance 0)

0.113

0.044

0.203

S1 (standard 1)

47.00

2.000

12.00

A1 (absorbance 1)

1.862

1.832

1.871

Table 2 Evaluation as Ammonium NH

4

LCK 303

LCK 304

LCK 305

RES (resolution)

--.--

-.---

--.--

S0 (standard 0)

02.50

0.020

01.30

A0 (absorbance 0)

0.110

0.047

0.198

S1 (standard 1)

60.00

2.500

15.00

A1 (absorbance 1)

1.845

1.783

1.788

Advertising