Appendix a modbus register information – Hach-Lange 1720E Low Range Turbidimeter User Manual

Page 45

Advertising
background image

41

Appendix A Modbus Register Information

Table 8 Sensor Modbus Registers

Group Name

Tag Name

Register

#

Data

Type

Length

R/W

Units (U)

Range

Measurements

Cal Gain

40013

float

2

R

none

0.5 to 2.0

Verification

PF Criteria

40062

Integer

1

R/W

%

5 to 10

Measurements

Turbidity

40001

Float

2

R

NTU

0/100

Diagnostics

Temperature

40005

Float

2

R

Deg C

Diagnostics

Dark Reading

40009

Float

2

R

NTU

0/100

Diagnostics

Raw Turbidity

40011

Float

2

R

NTU

0/100

Diagnostics

Lamp Voltage

40018

Float

2

R

Volts

0/5

Diagnostics

Lamp Current

40020

Float

2

R

Amps

0/2.75

Diagnostics

Plus 5V

40022

Float

2

R

Volts

0/5.5

Diagnostics

Voltage In

40024

Float

2

R

Volts

0/1443

Setup

Software Version

40015

Float

2

R

Setup

Bubble Rej

40017

Integer

1

R/W

On/Off

Setup

DataLog Interval

40026

Integer

1

R/W

Sec or Min

30 sec, 1 min,

5 min, 10 min,

15 min

Setup

Sensor Name

40027

String

6

R/W

Setup

Filter Size

40033

Integer

1

R/W

sec

no averaging,

6, 30, 60, 90

Setup

Sensor Ser Num

40036

String

6

R

12 digits

Setup

Output Mode

40042

Integer

1

R

0, 1, 2, 4

Setup

Set Resolution

40061

Integer

1

R/W

decimal places

4, 3, or 2