Turn the keypad sound off (optional), Standard operation, Prepare the stablcal standards – Hach-Lange 2100 N Laboratory Turbidimeter User Manual User Manual

Page 13: Calibration notes, Calibrate the turbidimeter with stablcal, Standards, Prepare the stablcal standards calibration notes

Advertising
background image

5. Push the power switch on the back of the instrument to turn the

instrument on.

Turn the keypad sound off (optional)

By default, the instrument makes an audible sound when a key is
pushed. To turn the keypad sound off:

1. Push and hold down the right arrow key for 3 seconds.
2. Use the arrow keys to select 00.
3. Push ENTER.
4. Use the arrow keys to select the sound option:

Option

Description

bP on

An audible sound is made when a key is pushed.

bP oF

No sound is made when a key is pushed.

5. Push ENTER.

Standard operation

Calibrate the turbidimeter with StablCal

®

Standards

Calibrate the turbidimeter before it is used for the first time using the
StablCal

®

sealed vial standards provided. As an alternative, calibration

can be done with recently prepared formazin standards. Refer to

Calibrate the turbidimeter with formazin standards

on page 32.

Calibrate the turbidimeter at least every 3 months or as specified by the
regulating authority when data is used for USEPA reporting.
The instrument is ready for calibration 60 minutes after start-up. Keep
the instrument on 24 hours a day if the instrument is used regularly.

Note: Unknown results may occur if standards other than the recommended
calibration points are used. The recommended calibration points (< 0.1, 20, 200,
1000 and 4000 NTU) provide the best calibration accuracy. Use of standards other
than StablCal, or user-prepared formazin, may result in less accurate calibrations.
The manufacturer cannot guarantee the performance of the instrument if calibrated
with co-polymer styrenedivinylbenzene beads or other suspensions.

Prepare the StablCal standards

When received and at intervals:

1. Clean the exterior surface of the StablCal vials with laboratory glass

cleaning detergent.

2. Rinse the vials with distilled or deionized water.
3. Dry the vials with a lint-free cloth.

Note: Never shake or invert the < 0.1 NTU standard. If the standard has been
mixed or shaken, do not move the vial for 15 minutes or more before using.
Note: Do not remove the caps from the sealed vials.

Make sure that the StablCal standards are at ambient instrument
temperature before use (and no greater than 40 °C (104 °F)).
Mix the standards before use:

1. Open the case lid. Remove the < 0.1 NTU standard from the plastic

case.

2. Leave the other standards in the case. Close the case lid.
3. Shake the case vigorously for at least 10 seconds.
4. Let the standards stand with no movement for 3–5 minutes before

use.

Calibration notes

• Make sure that the instrument is in the same ambient conditions as

where it is used.

• Make sure that the standards are at the same ambient temperature as

the instrument before use.

• Use only the provided silicone oil. This silicone oil has the same

refractive index as the vial glass and masks minor glass differences
and scratches.

• Store the oiling cloth in a plastic storage bag to keep the cloth clean.
• If power is lost during calibration, the new calibration data is lost and

the last calibration data is used. To exit a calibration and not save the
new values, push UNITS/Exit.

• In Calibration mode, automatic range and signal averaging on are

selected. When calibration is completed, all operational modes go
back to the last settings.

English 11