Delete calibration data, Flashing 9s, Flashing 0s – Hach-Lange 2100 AN Laboratory Turbidimeter User Manual User Manual

Page 55: Replacement parts and accessories, Delete calibration data flashing 9s flashing 0s

Advertising
background image

Table 9 Diagnostic codes (continued)

Code

Display

Description

23

Test results are shown.

Keyboard test

24

Test results are shown.

Memory test

Delete calibration data

To delete any calibration data entered by the user:

1. Turn off the instrument.
2. Push and hold CAL/Zero.
3. Turn on the instrument.

The CAL? light flashes. The instrument starts in Calibration mode.

4. Calibrate the instrument before use.

Flashing 9s

When manual ranging is selected, the display will flash all 9s when the
sample being measured is greater than the selected range.
When automatic ranging is selected, the display will flash 9s when the
sample is greater than the maximum range of the instrument. The
display will also flash 9s if Ratio is off and the measurement is greater
than 40 NTUs (268 nephelos or 9.8 EBCs). Turn Ratio on. Refer to

Measure over-range samples

on page 24.

Flashing 0s

When manual ranging is selected, the display will flash all 0s when the
sample measured is less than the selected range.
When automatic ranging is selected, the display will flash all 0s when the
measurement is less than the range of the instrument or a negative
value. Calibrate the instrument.

• When measuring color, absorbance or transmittance, set the zero

reference point again.

• When measuring absorbance, make sure that the reading is positive.

To measure samples with negative absorbance, set the analytical

zero using the sample with the greatest absorbance and read the
sample with the least absorbance. Record the reading as negative
absorbance.

Replacement parts and accessories

Note: Product and Article numbers may vary for some selling regions. Contact the
appropriate distributor or refer to the company website for contact information.

Replacement parts

Description

Quantity

Item no.

Calibration kit, StablCal

®

, sealed sample

cells (<0.1, 20, 200, 1000, 4000 and
7500 NTU)

1

2659505

Cover, sample cell compartment

1

4702500

Cover, lamp access

1

4703200

Cover, printer

1

4705400

Dust cover

1

4703000

Filter assembly, USEPA

1

3031200

Filter assembly, 455 nm

1

1999800

Fuse for 115 V operation, time-delay,
250 V, 1.6 A, UL/CSA approved

1

3030700

Fuse for 230 V operation, time-delay,
250 V, 1.6 A, IEC type, VDE approved

1

3030600

Gelex

®

secondary turbidity standardization

kit
(stray light standard and 0–2, 0–20, 0–200,
200-4000, and 4,000–10,000 NTU)

1

2589200

Lamp replacement kit

1

4708900

Oiling cloth

1

4707600

Power cord, North America, 115 VAC,
UL/CSA approved

1

1801000

English 53