Hach-Lange TSS PORTABLE Handheld Instrument Quick Guide User Manual

Quick guide, Tss portable handheld instrument, Turbidity measurement

Advertising
background image

1

DOC13.52.90141

Quick guide

TSS Portable Handheld Instrument

The instrument has two functions: turbidity measurement and suspended solids measurement. The appropriate
probe registered to the handheld instrument should be used for measurements. Refer to the safety information in
the manual before start-up.

An example of a turbidity measurement and a suspended solids measurement are given below. Refer to the
manual supplied for further information and details on the instrument.

Turbidity measurement

A standard curve C-TU is stored for the turbidity measurement. Calibration is not required.

1.

Connect probe to instrument.

2.

To switch the instrument on, press

ENTER/ON

for 2 seconds.

3.

Select

MEAS

and confirm with

ENTER/ON

.

4.

Place probe in measuring medium.

5.

Select

START

and confirm with

ENTER/ON

.

Note: The data is automatically saved every minute, max. 290 measurement points.

6.

To stop the measurement, select

MEAS

and confirm with

ENTER/ON

.

7.

Select

STOP

, confirm with

ENTER/ON

.

Delete the saved measurement points:

8.

Select

DATA

and confirm with

ENTER/ON

.

9.

Select

DELETE DATA

and confirm with

ENTER/ON

.

10. Select

YES

and confirm with

ENTER/ON

.

Suspended solids measurement

For the solids measurement, calibration is required. There are 4 curves available: C-DS1, C-DS2, C-DS3, C-DS4.
Each curve can be assigned to any measurement point. In the example, curve C-DS1 is selected.

1.

Connect the probe to the instrument.

2.

To switch on the instrument, press

ENTER/ON

for 2 seconds.

Select calibration curve

3.

Select

MENU

, confirm with

ENTER/ON

.

4.

Select

CALIBRATE

, confirm with

ENTER/ON

.

5.

Select

CURVE C-TU

and confirm with

ENTER/ON

.

Note: Curve C-TU is the default setting.

6.

Select calibration curve

C-DS1

with the navigation keys

UP/DOWN

and confirm with

ENTER/ON

.

The instrument reads the probe data and info text appears on the display.