Section 7 troubleshooting, 1 error messages, 2 informative messages – Hach-Lange TSS PORTABLE Handheld Instrument for Turbidity_Suspended Solids Manual User Manual

Page 31: 1 error messages 7.2 informative messages

Advertising
background image

Problem/display screen

Caused by

Action

Communication with probe

EEPROM error

Overflow area

Clean the probe

Offset error

Calibration error

Calibration insufficient –

Calibration insufficient +

Probe error

Battery is empty

Incorrect probe

Error time/date

Display

Caused by

Measure/note

Wait

Manufacturer standard

31

Section 7

Troubleshooting

7.1

Error messages

The probe is not connected to the
instrument

– Plug in the probe
– Check the connection to the probe
– Correct probe number

Values cannot be saved

Switch the instrument off and then back
on again

Concentration too high

– Probe is measuring in an unsuitable
medium
– It may be necessary to recalibrate

Probe is dirty

– Clean the probe
– May be triggered by inhomogeneous
measuring medium

Calibration point set incorrectly
(calibration point too close to the zero
point)

Recalibration required

No calibration available, calibration
points set incorrectly (upper and lower
calibration points transposed)

Recalibration required. (1-point default
calibration of DS curves not sufficiently
precise for measuring medium)

Further calibration point with lower
concentration is missing

Further calibration point with lower
concentration is required

Further calibration point with greater
concentration is missing

Further calibration point with greater
concentration required

Technical error in the probe

– Switch the instrument off and then
back on again
– If the message is repeated, send the
probe for repair

Battery not charged

Charge the battery fully

Probe has been replaced but not
re-registered or correctly registered on
instrument

– Connect the correct probe
– Use code 390 to release DS curve for
new probe and recalibrate

Time/date implausible

– Switch the instrument off and then
back on again
– Check the time/date set
– If required, change the clock
batteries

1

)

1

If the message is repeated, send the instrument for repair

7.2

Informative messages

Calibration is being recorded

Wait (probe in calibration medium)

Measurement value is outside the
standard (only for turbidity)

Value is above the DIN standard of
4000 FNU (NTU)