Standard operations – Hach-Lange BIOGAS User Manual User Manual

Page 35

Advertising
background image

35

Standard Operations

2.

Press 1 (New electrode).

3.

Press the validation key in the Function field

4.

Select the function of the electrode, refer to the

Table 3

.

5.

Press the validation key in the ID field.

6.

Select Other in the From field.

Note: The option From = Catalogue allows you
to create an electrode from a list of
preprogrammed electrodes.

7.

Enter the identification (ID) of the electrode, (up

to 16 characters).

8.

Press 1 twice.

9.

Select the electrode Type, for electrodes with

pH or mV functions:

Single pH - pH electrode (no reference

part),

Combined pH - pH electrode with a refe-

rence part (with or without temperature

sensor),

Single metal/redox - mV (i=0) function

electrode with a reference part,

Combined metal/redox - mV (i=0) func-

tion electrode with a reference part (with or

without temperature sensor).

Table 3 Electrode functions

Electrode type

Select function

pH single

pH

pH combined w/wo temp. sensor pH

pH

Single metal/redox

mV (i=0)

Combined metal/redox w/wo temp. sensor

mV (i=0)

ISE single

mV (i=0)

ISE Combined w/wo temp. sensor

mV (i=0)

Reference single

Reference

Temperature sensor

T°C

Ground metal

Ground

Double metal

mV (i>0)

Advertising