English 104

Press key 3 under OK to confirm the entered digestion term.

Enter the digestion temperature

The digestion temperature is determined by the boiling point of the
substance to be digested.
Press key 1 to adjust the digestion temperature in 5 °C increments over
a temperature range of 40 °C to 150 °C or 170 °C.
Press key 3 under OK to confirm the entered digestion temperature.

Influence of the acid concentration on the temperature

reached in the sample containers

Note: The programmed digestion temperature for the digestion of
aqueous solutions with diluted acids and lyes (boiling points of 100 to 110
°C) is the temperature that is reached in the reaction tubes or cuvettes
during the digestion process. The temperature in the thermostat block
may be up to 20 °C higher!
For the digestion of semi-concentrated to concentrated acids (boiling

points of 150 °C and above), the temperatures reached in the reaction
tubes or cuvettes are higher than the programmed digestion
temperature.
In this instance, the temperatures in the reaction tubes or cuvettes are
equivalent to the temperatures in the thermostat block.

Enter the digestion time

Press key 1 to adjust the digestion time (time) in 5 minute increments
over a range of 5 to 240 minutes.
Press key 3 under OK to confirm the entered digestion time.

Position (first,, 2. position etc.) of the user-specific

programs in the main menu

Press key 1 to specify the required position (1–9) on the display.
Press key 3 under OK to confirm the entered position.

Open the user-specific digestion programs

Press key 4 under the arrow symbol in the main menu to show
additional menu items:
Press the key under the relevant digestion term to select the
corresponding digestion program.
Note: This menu only opens if user-specific digestion programs have
been programmed.

Variation (VAR) of user-specific digestion programs

User-specific digestion programs are varied using the VAR menu.
Press key 2 under VAR in the main menu.
The Digestion term, Temperature, Digestion time and Position
parameters can be varied by the user.

Select the digestion program to be varied

Press key 1 under the arrow symbol in the main menu to show
additional menu items.
Press the keys under the relevant digestion term to select the digestion
program to be varied.

Selected

digestiontem

perature

(°C)

Temperature (°C) in the reaction tubes/

cuvettes for digestions of:

Thermostat

block

temperature

(°C)

Aqueous solutions

with diluted acids and

lyes

Semi-concentrated to

concentrated acids

40

40

41

41

60

60

62

62

80

80

84

84

100

100

110

110

120

120

134

134

130

130

146

146

135

135

152

152

140

140

158

158

150

150

170

170