4

7470 Series

Installation Instructions

1) Inspect meter for damage that may have occurred during shipping. Report any

damage to the container to the freight carrier immediately.

This is important information. Read it carefully before

beginning work.

2) Make sure your pressure, temperature, fluid and other requirements are

compatible with the meter including o-rings.

3) Select a suitable location for installation to prevent excess stress on the meter

which may result from:

a) Misaligned pipe.

b) The weight of related plumbing.

c) "Water Hammer" which is most likely to occur when flow is suddenly stopped

as with quick closing solenoid operated valves. (If necessary, a surge chamber

should be installed. This will also be useful in pressure start-up situations.)

d) Thermal expansion of liquid in a stagnated or valve isolated system.

e) Instantaneous pressurization which will stress the meter and could result in

tube failure.

NOTE: In closed thermal transfer or cooling systems, install the meter in the cool

side of the line to minimize meter expansion and contraction and possible fluid

leaks at the threaded connections.

4) Handle the meter carefully during installation.

a) Use an appropriate amount of teflon tape on external pipe threads before

making connections. Do not use paste or stick type thread sealing products.

5) Install the meter vertically with the inlet port at the bottom.

Maximum Non-Shock

Pressure and Temperature O-Ring Temperature

Viton and Kalrez are registered trademarks of DuPont Dow Elastomers.

Temp.

Pressure

Ambient temperature

33° F - 125° F

EPR

225 °F

Buna-N
Viton
Kalrez

275 °F
350 °F
400 °F

250 psig

(31W-43W)

250°F

(gas)

200°F

(liquid)

Maximum

Temperature

O-Ring

Material

b) Over tightening of plastic connections may result in fitting damage.

6) Connections: Fittings are fully rotatable by loosening the fitting retainer screws

during installation. Make sure fitting retainer screws are tightening after adjustment

of fitting.

7) Meters with stainless steel fittings will support several feet of pipe as long as

significant vibration or stress resulting from misaligned pipe are not factors.

8) Meters with plastic fittings must be installed so that fittings are not made to

support any part of the associated plumbing. In addition, meter frame should be

fastened to bulkhead, panel or column.

9) Meters used in gas service should have suitable valves plumbed in at the inlet

and outlet of the meter. These valves should be no more than 1-1/2 pipe diameters

from the meter ports. The valve at the outlet should be used to create back

pressure as required to prevent float bounce. It should be set initially and then left

alone. The inlet valve should be used for throttling purposes. Depending on the

installation, valves may not be essential, but they are most useful in many

installations. Remember: To get a correct reading of flow in gas service, it is

necessary to know the pressure right at the outlet of the meter (before the valve).

200 psig

(51W-67W)

110°F

(liquid)

150 psig

(31W-94W)

316LSS

PVC

150 psig

(81W-94W)

CASE ASSEMBLY

SIDE PLATE

SIDE PLATE SCREW

ALARM SWITCH HOUSING

ALARM SWITCH HOUSING SET SCREW

ALARM SWITCH

ALARM GUIDE ROD NUT

ELECTRICAL STRAIN CONNECTOR

JUNCTION BOX O-RING

JUNCTION BOX WITH COVER

JUNCTION BOX MOUNTING SCREW

4X

2X

2X

2X

2X

2X

ALARM GUIDE ROD

ALARM GUIDE ROD BRACKET

ALARM GUIDE ROD BRACKET SCREW

7470 series asseMbly

with alarM
parts list:

1. Case Assembly
2. Side Plate
3. Side Plate Screw
4. Junction Box O-Ring
5. Junction Box With Cover
6. Junction Box Mounting Screw
7. Alarm Guide Rod Bracket
8. Alarm Guide Rod Bracket Screw
9. Alarm Guide Rod Nut
10. Alarm Guide Rod
11. Alarm Switch Housing Set Screw
12. Alarm Switch Housing
13. Alarm Switch
14. Electrical Strain Connector

latching reed switch

All 7910 Series flowmeters may be fitted with one or two latching reed
switches.
The switch assembly is mounted on the guide rod. The switch can be
positioned to trip at any point on the scale.
The switch is a reed type and uses a biasing magnet to give it the
latching feature. The float contains hermetically sealed magnet(s), so
when the float comes in proximity to the switch it closes and remains
closed (latched) when the float moves below the switch it resets itself.
Contact King Instrument Company for multiple switch options.

latching reed switch-electrical
speciFications

TYPE:

SPDT

/

Latching

MAXIMUM CONTACT VOLTAGE:

100V DC

MAXIMUM CONTACT CURRENT:

0.20 A DC

MAXIMUM CONTACT POWER:

4 Watts DC

BREAKDOWN VOLTAGE:

200V DC

STANDARD PULL-IN RANGE:

15-40 Ampere turns

INITIAL CONTACT RESISTANCE:

0.150 Ohm

connections - intrinsically saFe wiring

SWITCH ISOLATER 3 CONDUCTOR, 22Awg, 2’ Long
1) White - N.O. switch output 1
2) Red - common
3) Black - N.C. switch output 2

switch isolator option:

Latching reed switches can be used as stand alone devices, or may be
connected to a switch isolator for intrinsically safe applications. The
purpose of the switch isolator is to supply electrical signals between
safe and hazardous areas in either direction while limiting the amount
of energy that can be transferred even under fault conditions. Switch
isolators are available with 220VAC, 110VAC or 24VDC supply voltage
requirements, contain single pole double throw (SPDT) relays, and are
DIN rail mountable. See switch isolator specifications for electrical con-
nections and further details.

Float types and orientations

GV FLOAT

GS FLOAT

LP FLOAT

SL FLOAT

Pressure and temperature ratings are based on a study of the engineering data for

particular materials used in construction and on the design of individual models.

This information is supplemented by destructive test results. Meters with stainless

enclosures must never be operated without shields securely in place. Meters

exposed to difficult environments such as those created by certain chemicals,

excessive vibration or other stress inducing factors could fail at or below the

suggested maximums. Never operate meters above pressure and temperature

maximums. It is strongly recommended that all meter installations utilize an

appropriate pressure relief valve and/or rupture disc. The pressure settings and

locations of these devices should be such that meters cannot be over pressurized.

Meter failure could result in damage to equipment and serious personal injury.

Always use suitable safety gear, including OSHA approved eye protection when

working around meters in service. We are happy to pass along chemical

compatibility information that has been published by the manufacturer's of raw

materials used in our products; however, this information should not be construed

as a recommendation made by King Instrument Company, Inc. for a specific

application.

Carefully remove the flowmeter from piping system. Remove the 4 screws on each

side holding the side plates on. Remove the side plates. Carefully remove the glass

meter tube from the end fittings. Be sure to not let the internals fall out. Use caution

when removing the glass meter tube. Do not allow float to fall out. Float damage

may result in inaccuracy. All necessary instrument components are now fully

accessible for cleaning with a bottle brush and appropriate mild soap solution*.

Before the meter is reassembled, inspect all parts for damage. O-rings should be

replaced during meter maintenance and cleaning.

To reassemble, install the glass meter tube back onto the end fittings. Reinstall the

side plates. Tighten the 4 screws on each side. Reinstall the instrument into the

plumbing system after removing the old teflon tape (with a wire brush) and

replacing with fresh teflon tape.

*Do not use cleaning agents that will damage float, tube or o-rings.

Meters should be cleaned with a mild soap solution. This will be an effective

cleaner of rust stains. Caution must be used so that materials of construction are

not damaged by cleaning solutions. Hard water deposits can be removed with 5%

acetic acid solution (vinegar).

7470 meters that require repair should be sent to the factory. Please call for a

Return Merchandise Authorization (RMA) number and return instructions.

-O-rings should be replaced if meter is disassembled after it has been in service.

-Serious property damage and great personal injury could occur as the result of a

meter misused or used in an unsuitable application.

CAUTION:

CLEANING:

REPAIR:

WARNING:

FLOAT TYPES AND

ORIENTATIONS:

LATCHING REED SWITCH

All 7910 Series flowmeters may be fitted with one or two latching reed switches.

The switch assembly is mounted on the guide rod. The switch can be positioned to

trip at any point on the scale.

The switch is a reed type and uses a biasing magnet to give it the latching feature.

The float contains hermetically sealed magnet(s), so when the float comes in

proximity to the switch it closes and remains closed (latched) when the float moves

below the switch it resets itself. Contact King Instrument Company for multiple

switch options.

LATCHING REED SWITCH-ELECTRICAL

SPECIFICATIONS

TYPE:

MAXIMUM CONTACT VOLTAGE:
MAXIMUM CONTACT CURRENT:
MAXIMUM CONTACT POWER:
BREAKDOWN VOLTAGE:

SPDT / Latching

100V DC

0.20 A DC
4 Watts DC
200V DC

STANDARD PULL-IN RANGE:

15-40 Ampere turns

INITIAL CONTACT RESISTANCE:

0.150 Ohm

SWITCH ISOLATOR OPTION:

Latching reed switches can be used as stand alone devices, or may be connected

to a switch isolator for intrinsically safe applications. The purpose of the switch

isolator is to supply electrical signals between safe and hazardous areas in either

direction while limiting the amount of energy that can be transferred even under

fault conditions. Switch isolators are available with 220VAC, 110VAC or 24VDC

supply voltage requirements, contain single pole double throw (SPDT) relays, and

are DIN rail mountable. See switch isolator specifications for electrical connections

and further details.

CONNECTIONS - INTRINSICALLY SAFE WIRING

SWITCH ISOLATER 3 CONDUCTOR, 22Awg, 2' Long

1) White - N.O. switch output 1

2) Red - common
3) Black - N.C. switch output 2

READ

HERE

GV FLOAT

Highest immunity to

viscous fluids with

medium meter

capacity.

SL FLOAT

Maximum flow

meter capacity with

limited viscosity

immunity

READ

HERE

LP FLOAT

Maximum flow

meter capacity with

limited viscosity

immunity

READ

HERE
READ

HERE

GS FLOAT

Maximum flow

meter capacity with

limited viscosity

immunity