7330 series, Installation instructions – KING 7330 Series User Manual

Page 4

Advertising
background image

(714) 891-0008 • www.kinginstrumentco.com

When it comes to flow...we’re instrumental.

4

7330 Series

Installation Instructions

Installation Instructions

7330 series asseMbly with latching reed switch
parts list:

1. End Fitting Union Nut
2. End Fitting
3. End Fitting O-Ring
4. Float Stop/ Extension Assembly
5. Float
6. Polysulfone Meter Tube Assembly
7. Alarm Switch
8. Alarm Switch Housing
9. Alarm Set Screw O-Ring
10. Alarm Set Screw
11. Alarm Switch Housing Screw

7330 series latching reed switch

All 7330 Series flowmeters may be fitted with one or two latching
reed switches.
The switch assembly is mounted on the meter tube. The switch can
be positioned to trip at any point on the scale.
The switch is a reed type and uses a biasing magnet to give it the
latching feature. The float contains hermetically sealed magnet(s),
so when the float comes in proximity to the switch it closes and
remains closed (latched) when the float moves below the switch it
resets itself. Contact King Instrument Company for multiple switch
options.

latching reed switch electrical speciFications

type:

spdt / latching

MAXIMUM CONTACT VOLTAGE:

100V DC

MAXIMUM CONTACT CURRENT:

0.20 A DC

MAXIMUM CONTACT POWER:

4 Watts DC

BREAKDOWN VOLTAGE:

200V DC

STANDARD PULL-IN RANGE:

15-40 Ampere turns

INITIAL CONTACT RESISTANCE:

0.150 Ohm

connections - intrinsically saFe wiring

SWITCH ISOLATER 3 CONDUCTOR, 22Awg, 2’ Long
1) White - N.O. switch output 1
2) Red - common
3) Black - N.C. switch output 2

switch isolator option:

Latching reed switches can be used as stand alone devices, or may
be connected to a switch isolator for intrinsically safe applications.
The purpose of the switch isolator is to supply electrical signals
between safe and hazardous areas in either direction while limit-
ing the amount of energy that can be transferred even under fault
conditions. Switch isolators are available with 220 VAC, 110 VAC or
24 VDC supply voltage requirements, contain single pole double
throw (SPDT) relays, and are DIN rail mountable. See switch isolator
specifications for electrical connections and further details.

12700 Pala Drive, Garden Grove, CA 92841

www.kinginstrumentco.com / (714) 891-0008

Meters are warranted against defects in materials and workmanship

to the original user for a period of thirteen (13) months from the date

of factory shipment, provided the meter is installed, operated and

maintained in accordance with King Instrument Company's

instructions and recommendations.

This warranty does not apply if failure is caused or contributed to by

any of the following: improper handling, improper storage, abuse,

unsuitable application of the product, lack of reasonable and

necessary maintenance, use exceeding suggested pressure and

temperature maximums, improper packaging for return, or repairs

made or attempted to be made by anyone other than King Instrument

Company, Inc.

KING INSTRUMENT COMPANY, INC. MAKES NO

WARRANTY AS TO THE FITNESS OF ITS PRODUCTS

FOR SPECIFIC APPLICATIONS.

This warranty is valid for the original end-user only and does not

apply to products that have been damaged or modified. This

warranty is non-transferrable and is limited to replacement or repair.

The liability of King Instrument Company arising out of its supply of

the products, or their use, shall not in any case exceed the cost of

correcting defects in the products as set forth above.

THIS WARRANTY IS A LIMITED WARRANTY AND

SHALL BE IN LIEU OF ANY OTHER WARRANTIES,

EXPRESSED OR IMPLIED, INCLUDING BUT NOT

LIMITED TO ANY IMPLIED WARRANTY OR

MERCHANTABILITY OR FITNESS FOR A PARTICULAR

PURPOSE. THERE ARE NO OTHER WARRANTIES

WHICH EXIST BEYOND THE DESCRIPTION OR FACE

HEREOF.

IN NO EVENT SHALL KING INSTRUMENT COMPANY BE

LIABLE FOR LOSS OF PROFITS, INDIRECT,

CONSEQUENTIAL OR INCIDENTAL DAMAGES.

Products should be returned, prepaid, to King Instrument Company,

Inc. with proof of purchase. Call factory for Return Merchandise

Authorization (RMA) number and return instructions.

2013A

All 7330 Series flowmeters may be fitted with one or two latching reed switches.

The switch is a reed type and uses a biasing magnet to give it the latching

feature. The float contains hermetically sealed magnet(s), so when the float

comes in proximity to the switch it closes and remains closed (latched) when

the float moves below the switch it resets itself. Contact King Instrument

Company for multiple switch options.

LATCHING REED SWITCH ELECTRICAL

SPECIFICATIONS

TYPE:

MAXIMUM CONTACT VOLTAGE:
MAXIMUM CONTACT CURRENT:
MAXIMUM CONTACT POWER:

SPDT / Latching

100V DC

0.20 A DC

4 Watts DC

BREAKDOWN VOLTAGE:

200V DC

STANDARD PULL-IN RANGE:

15-40 Ampere turns

INITIAL CONTACT RESISTANCE:

0.150 Ohm

SWITCH ISOLATER 3 CONDUCTOR, 22Awg, 2' Long

END FITTING

END FITTING O-RING

FLOAT

EXTENSION ASSEMBLY

ALARM SWITCH HOUSING

FLOAT STOP/

ALARM SWITCH

ALARM SET SCREW

ALARM SWITCH HOUSING

UNION NUT

END FITTING

POLYSULFONE

The switch assembly is mounted on the meter tube. The switch can be

positioned to trip at any point on the scale.

CONNECTIONS - INTRINSICALLY SAFE WIRING

1) White - N.O. switch output 1

2) Red - common
3) Black - N.C. switch output 2

SWITCH ISOLATOR OPTION:

Latching reed switches can be used as stand alone devices, or may be

connected to a switch isolator for intrinsically safe applications. The purpose of

the switch isolator is to supply electrical signals between safe and hazardous

areas in either direction while limiting the amount of energy that can be

transferred even under fault conditions. Switch isolators are available with

220VAC, 110VAC or 24VDC supply voltage requirements, contain single pole

double throw (SPDT) relays, and are DIN rail mountable. See switch isolator

specifications for electrical connections and further details.

METER TUBE ASSEMBLY

ALARM SET SCREW

O-RING

SCREW