A&D UA-704 User Manual

Page 9

Advertising
background image

TROUBLESHOOTING

SPECIFICATIONS

Problem

Display is blank,
even when the
power is turned
on.

The device will
not measure.
Readings are too
high or too low.

Other

Probable cause

Battery is drained.

Battery terminals are
not in the correct
position.

The cuff is not
fastened properly.

You moved your arm
or body during the
measurement.

The cuff position is
not correct.

You are wearing the
wrong cuff size.

The blood pressure
reading is different
from that measured
at a clinic or doctor’s
office.

Cuff will not inflate.

NOTE: If the actions described above do not solve the problem, contact
A&D Medical or Auto Control Medical at the numbers listed at the
bottom of these pages. Do not attempt to repair the device yourself.

LifeSource Health Line (Toll-Free): 1-888-726-9966

In Canada (Toll-Free): 1-800-463-5414

E-14

E-15

Model

. . . . . . . . . . . . . . . . . . .UA-704

Type

. . . . . . . . . . . . . . . . . . . . .Oscillometric

Display

. . . . . . . . . . . . . . . . . .Digital, 10-mm character height

Pressure/pulse displayed
alternately

Measurement range

. . . . . . . . .Pressure: 20 mmHg to 280 mmHg

Pulse: 40 pulses to 200
pulses/minute

Accuracy

. . . . . . . . . . . . . . . . .Pressure: +3 mmHg or 2%,

whichever is greater
Pulse: +5%

Pressurization

. . . . . . . . . . . .Manually by inflator bulb

Depressurization

. . . . . . . . . .Constant-air release-valve system

Deflation

. . . . . . . . . . . . . . . . .Hand-held push button

Power source

. . . . . . . . . . . . .1 AA battery (1.5 volt, included)

Battery life

. . . . . . . . . . . . . . .Approximately 2,000

measurements

Operating environment

. . . .50ºF to 104ºF (10ºC to 40ºC)

Less than 85% relative humidity

Storage environment

. . . . . .-4ºF to 140ºF (-10ºC to 60ºC)

Less than 95% relative humidity

Dimensions

. . . . . . . . . . . . . . .Length: 3.9”(100 mm)

Width: 1.2”(31 mm)
Height: 2.0” (51 mm)

Weight

. . . . . . . . . . . . . . . . . . .2.6 oz. (73 g) without battery

Blood pressure measurements determined by the UA-704 are equivalent to those
obtained by a trained observer using the cuff/stethoscope auscultation method
within the limits prescribed by the ANSI/AAMI SP-10 standard for electronic or
automated sphygmomanometers.

Corrective action

Replace with a new battery.

Place the battery with
negative and positive
terminals matching those
indicated in the battery
compartment.

Fasten the cuff correctly.

Make sure you remain very
still and quiet during the
measurement.

Adjust the cuff position.
Raise your hand so that the
cuff is at the same level as
your heart.

See Pg. E-5 “Selecting the
Correct Cuff.”

See Pg. E-12 “Keys to
Successful Monitoring.”

Remove the battery.
Place back in correctly, and
try the measurement again.

Check air tube connection
inside battery cover.