ELECTRONICS INC., 3183-G Airway Ave., Costa Mesa, CA 92626, USA • Tel 714-434-6131 • www.laurels.com 2

Specifications

Analog Input

Range

Resolution

Input Ohms

Error at 25°C

200.00 mV

2.0000 V
20.000 V
200.00 V

10 µV

100 µV

1 mV

10 mV

1 MΩ

± 0.1% of FS

Voltage Ranges

0% to 100% of FS,

10 Hz to 10 kHz

600.0 V*

100 mV

1 MΩ

± 0.3% of FS

2.0000 mA
20.000 mA
200.00 mA

0.1 µA

1 µA

10 µA

100 Ω

10 Ω

1 Ω

± 0.1% of FS

Current Ranges

0% to 100% of FS,

10 Hz to 10 kHz

5.000 A

1 mA

0.01 Ω

± 0.4% of FS

Analog Output (standard)

Output Levels
Compliance, 4-20 mA
Compliance, 0-10V
Output Resolution
Output Accuracy
Output Isolation

0-20 mA or 0-10 Vdc (selectable)
10V ( 0-500Ω load )
2 mA ( 5 kΩ load )
16 bits (65,536 steps)
0.02% of output span plus conversion accuracy
250V rms working, 2.3 kV rms per 1 minute test

Serial Data Output (standard)

Type
Data Rates
Output Isolation
Serial Protocols
Modbus Compliance
Digital Addresses

10/100Base-T Ethernet per IEEE 802.3
300, 600, 1200, 2400, 4800, 9600, 19200 baud
250V rms working, 2.3 kV rms per 1 min test
Modbus TCP, Modbus RTU, Modbus ASCII, Laurel ASCII
Modbus over Serial Line Specification V1.0 (2002)
247 for Modbus, 31 for Laurel ASCII

Dual Relay Output (standard)

Relay Type
Load Rating

Two solid state relays, SPST, normally open, Form A
120 mA at 140 Vac or 180 Vdc

Power Input

Standard Power
Low Power Option
Power Frequency
Power Isolation
Power Consumption

85-264 Vac or 90-300 Vdc (DC operation not ETL certified)
10-48 Vdc or 12-32 Vac
DC or 47-63 Hz
250V rms working, 2.3 kV rms per 1 min test
2 Watts

Mechanical

Dimensions
Mounting
Signal & Power

129 x 104 x 22.5 mm case
35 mm rail per DIN EN 50022
Detachable screw-clamp connectors

Environmental

Operating Temperature
Storage Temperature
Relative Humidity
Cooling Required

0°C to 55°C
-40°C to 85°C
95% at 40°C, non-condensing
Mount transmitters with ventilation holes at top and bottom.
Leave 6 mm (1/4") between transmitters, or force air with a fan.

* Range ETL certified to 300.0 V. For purposes of accuracy calculation, the full scale range for the 200V