Calibration, Calibrating the cim100 – Liquid Controls CIM100 User Manual

Page 14

Advertising
background image

calibrating the cIM100

1...Calibrate.the.CIM100.with.a.volumetric.prover..
2...Determine.the.C-factor.
3...Enter.the.C-factor.into.the.CIM100.C-factor.field.

1. calibrate the cIM100 with a volumetric prover.

1... Remove.the.CIM100.cover.
2.. Scroll.to.the.Decimal.option.and.set.the.decimal.display.to.hundreths.

(0.00)

3.. Scroll.to.the.C-factor.option.and.enter.the.approximate.C-factor..Enter.

1.377410.3.for.gallons..Enter.5.208333.3.for.litres.

4.. Run.a.delivery.through.the.CIM100.and.into.the.volumetric.prover.at.

maximum.or.operating.flow.rate..

5.. Write.down.the.volume.delivered.into.the.volumetric.prover.and.the.

volume.displayed.on.the.CIM100..You.will.need.both.volumes.to.

determine.the.C-factor.

2. determine the c-factor.

1.. Calculate.the.%.Error.
2.. Calculate.the.K-Factor.
3.. Calculate.the.C-Factor.

x10

-1

Value to the right

of decimal point determines

the negative power

x10

-1

= One decimal place to the left

Real value move decimal

point to the left at x10

-x

.

x = number at the far right

Sets decimal point on the

display counter to hundreths

calIbratIon

K-factor

K-factor = (1 - )Approximate K-factor

ExamplE

191.997696 = (1 -

)192

Approximate K-factor for the CIM100 is

726 pulses per gallons

192 pulses per litre

Approximate C-factor for the CIM100 is

726 = 0.001377410

192 = 0.005208333

c-factor

C-factor =

X

K-factor

ExamplE

0.005208396 =

1

191.997696

% Error

Prover Volume - Metered Volume

Prover Volume

% Error =

100 - 99.88

100

.0012 =

ExamplE

ThE C-FACTOR IS ThE INVERSE OF

ThE K-FACTOR If the K-factor is 100, the

C-factor is ¹⁄₁₀₀ or 0.01.

ThE K-FACTOR IS ThE NuMbER

OF PuLSES PER uNIT OF

MEASuREMENTS

A PuLSE OCCuRS EACh TIME A

TuRbINE bLADE PASSES ThE PICKuP

COIL

The series of equations apply the CIM100

totalizer volume and the prover volume

recorded during the proving run in order to

calculate the C-factor.

For the purposes of calibrating the

CIM100:

.0012

100

% Error

100