Calibration, Sensitivity, Calibration factor – Liquid Controls IT200N User Manual

Page 6: Formula 2

Advertising
background image

Sponsler, Inc.

IT200N Remote Rate Indicator

pg. 6

DOC#: MN-200N-D.DOC

Frequency =

K-Factor x Flowrate

60





CALIBRATION



Sensitivity

The sensitivity adjust R1 is located on the IT200I (Input and Digital) P.C.B. The amplitude of the signal
generated by the turbine is proportional to the rate of flow; therefore, sensitivity should be adjusted at
the lowest usable flow rate. Rotate R1 completely counterclockwise then slowly rotate R1 clockwise until
the display correctly responds then increase R1 slightly clockwise. The nominal R1 position is with the
arrow indicating 11 o’clock.


FIGURE 2















Calibration Factor

The calibration factor is derived from the turbine’s K-Factor (Pulses per gallon or other desired
engineering unit).

C.F. =

K-Factor (Gallon)

Engineering Units


EXAMPLE 1: K-Factor = 250 pulses per gallon

Engineering Units = gallons

∴ C.F. = 250/1 = 250

On the Factoring P.C.B.:

Set S1@0, S2@5, S3@2, S4@0, S5@0 (S5 is most significant digit,

∴ CF entered is

00250)


The Calibration Factor is initially loaded @ ‘POWER ON’ or by activating RESET.

The electrical accuracy can be verified by injecting a stable frequency @ TB2-1,2 on the Mounting
P.C.B. and incorporating the following formula:

Rate =

F x 60

C.F.


Where F = Frequency in Hz (

)

Formula 1

Formula 2