Calibration ‘0-1’ function – Liquid Controls IT150N User Manual

Page 7

Advertising
background image

Sponsler, Inc.

IT150N Remote Totalizer

pg

7

DOC# MN-150N-B

41,381,600 x 1.2566

100




EXAMPLE #2: F = 500 Hz T = 1 minute (60 sec) C.F. = .4000

Total Displayed =

500 x 60 x .4000

=

30,000 x .400

=

12000

= 120 in 1 minute

100

100

100


FIGURE 3















Calibration ‘0-1’ Function

The ‘0-1’ function provides enhanced accuracy when totalization encompasses a large quantity or an
extended period of time such as SCF produced in a 24 hour period.

The ‘0-1’ function should be incorporated only when both the conditions listed below are met:

1) C.F.’s 1

st

digit right of decimal is 1

2) C.F.’s 5

th

digit right of decimal is not 0


Example #3

Assume a turbine has a K-Factor of 79.58 pulses per SCF and the customer product
demand is 520,000 SCF a day


C.F.

= 1/79.58

= .0125659 = .12566

÷10 Note: Both conditions are met


Without the ‘0-1’ function:

Set S6 #2 ‘ON’ (

↓ Position) for ÷ 10 (moves decimal point right 1 place)

Set S2@1, S3@2, S4@5, S5@7 (.1257 rounding fourth digit)

Set S1 in ‘0’ Position (0.1257) on the IT150I P.C.B.

A usage of 520,000 SCF = 41,381,600 total pulses (520,000 x 79.58) and using the C.F. of S1-S6
the displayed quantity is 520,166 SCF ( )

rather than 520,000 for a difference of 166 SCF.

With the ‘0-1’ function:

Set S6 #3 ‘ON’ (

↓ Position) for ÷ 100 (moves decimal right 2 places)

Set S2@2, S3@5, S4@6, S5@6 (.2566)

Set S1 in ‘1’ Position (1.2566) on the IT150I P.C.B.


As stated above the 24 hr usage is 520,000 SCF. The displayed quantity is now 520,001 SCF

( ) for a difference of 1 SCF.

41,381,600 x .1257

10