PHOENIX 1400

LTPHO1400 REV. 005 04/01/2013


PARAMETER

MIN

TYP

MAX

UNITS

Wavelength

Mode Hop Free Tuning Range

1515

1565

nm

Wavelength Set Point Resolution

0.01

pm

Absolute Accuracy

1

±1.5

pm

Wavemeter Accuracy

2

±0.5

pm

Wavemeter Linearity

±50

fm

Stability

3

±50

pm

Tuning

Tuning Rate

1

100

nm/s

Deviation from Linearity

4

±1.5

±2

GHz

Power

Range

6

10

mW

Accuracy

±5

%

Flatness

5

±7

±10

%

Ripple

5

±2

%

Noise

Spectral line width

6,7

1.5

MHz

Side mode suppression ratio (nearest mode)

6

43

50

dB

Relative intensity noise

6

-152

-145

dB/Hz

Spontaneous Emission Ratio

6,8

-50

-43

dBc/nm

Inputs

Auxiliary Detectors (SM FC/APC)
Min detectable
Max detectable
Damage threshold

0.02

10

0.03

12
50

0.05

15

mW

Electrical Trigger (BNC)

“Low” voltage

“High” voltage

-0.5

+2.0

0

+3.3

+0.8
+3.8

V

Outputs

Optical Power

SM FC/APC

Electrical Sweep and Retrace Triggers (BNC)
+1 M

load

+50 M

load


+2.5
+1.7

V

All measurements assume warm-up time of 1.5 hrs.
1

Determined by internal NIST-traceable gas cell reference.

2

Measured after wavelength auto-calibration.

3

For 24 hr period within ± 5 C after laser calibration.

4

Over full wavelength range at 10 nm/s.

5

At 10 mW tuning at 10 nm/s over full wavelength range.

6

Measured with laser set to 1540 nm with10 mW output power.

7

Phase noise distribution full width at half maximum at center wavelength integrated over 1 ms. Measurements indicate that the intrinsic laser
linewidth (the limit at which the integration time goes to zero) is less than 100 kHz.

8

Measured with optical spectrum analyzer set at 1 nm resolution bandwidth.


CLASS 1 LASER PRODUCT