Ova 5000 lp – Luna OVA 5000 LP User Manual

Page 2

Advertising
background image

Optical Vector Analyzer

TM

is a trademark of Luna Innovations

Incorporated.

©Luna Innovations Incorporated, Roanoke, VA
All rights reserved.

OVA 5000 LP

LT0VA5000LP REV. 005 02/13/2014


PARAMETER

FAST MODE*

AVERAGING MODE*

UNITS

Wavelength Range:

1270-1340 or 1525-1610

nm

Wavelength:

Standard Resolution

1.6

1.6

pm

Accuracy

1

±1.5

±1.5

pm

Repeatability

±0.1

±0.1

pm

Insertion Loss Characteristics:

Dynamic range

2

60

80

dB

Ripple

±0.05

±0.01

dB

Resolution

±0.05

±0.002

dB

Insertion loss accuracy

±0.1

±0.05

dB

Return loss accuracy

±0.2

±0.1

dB

PMD:

Range

3

6

6

ns

Accuracy

– First Order

±0.03 (100pm steps)

±0.15 (30pm steps)

±0.08

ps

Accuracy

– Second Order

±10

±2

ps

2

Loss range

2

40

50

dB

PDL:

Extinction ratio

40

50

dB

Accuracy

±0.05

±0.03

dB

Measurement Timing:

Laser sweep rate

70

70

nm/s

All-parameter measurement rate

4

30

30

ms/nm

Fully specified measurement time

5

12

55

s

Real-time mode update rate

6

1

NA

Hz

Maximum Device Length (including leads):

Transmission

150

150

meters

Reflection

75

75

meters

Physical

Weight (Processor not Included)

16.24

35.8

kg

lbs

Case Size (W X D X H)

473 X 420 X 206

18.62 X 16.54 X 8.08

mm

inches

CLASS 1 LASER PRODUCT

1.

Accuracy maintained by an internal NIST-traceable HCN gas cell.

2.

8

0 and 50 dB dynamic ranges in ‘Averaging Mode’ for IL and

PMD are with the “High Dynamic Range Averaging” option
installed and enabled.

3.

Specifies the total device impulse-response duration that may be
captured.

4.

Rate calculated from combined laser sweep and analysis time per
scan.

5.

Measurement with full specification (see note 2) over Fast Mode:
40 nm range, and Averaging Mode: 2.5 nm range. Excludes
calibration time.

6.

For 2.5 nm scan range.

Fast Mode: no averaged calibration scans, 4 averaged
measurement scans, 30 pm resolution bandwidth, 8 m device
length (accuracies verified using NIST certified artifacts except for
IL). High dynamic range option enabled.
Averaging Mode: 4 averaged calibration scans, 64 averaged
measurement scans, 30 pm resolution bandwidth, 8 m device
length (accuracies verified using NIST certified artifacts except for
IL). High dynamic range option enabled.
*results are typical

Advertising