PARAMETER

SPECIFICATION

UNITS

Performance

Model

A10

A50

Maximum Sensing Length

10

50

meters

Acquisition Rate

1

5

2.5

Hz

Minimum Sensor Spacing

0.4

mm

Minimum Gage Length

2

1

mm

Wavelength Accuracy

3

1.5

pm

Strain:

Range

± 13,000

µStrain

Single-scan repeatability

1,4,5

± 2

µStrain

Temperature:

Range

6

-50 to 300

°C

Single-scan repeatability

1,4,5

± 0.2

°C

Physical

Dimensions

14 x 12.5 x 6.75

(36 x 32 x 17)

in (cm)

Weight

17.5

(8)

lb (kg)

Power Consumption

50

W

CLASS 1 LASER PRODUCT

1

For the default measurement range of ±1,250 µStrain or ±200 °C; see Page 3 for detailed timing information

2

Minimum gage length is achievable using the largest measurement range, having a single-scan repeatability of ±17.0 µStrain or
±2.0 °C

3

Accuracy maintained by an internal NIST-traceable HCN gas cell

4

Temperature and strain measurements are calculated from the spectral shift of scattered laser light. Using the default conversion
coefficients of 1 GHz = 0.8 °C = 6.58 µStrain, the accuracy of temperature and strain are 0.15°C and 1.25 µStrain. [Othonos and
K Kalli, Fiber Bragg Gratings (Artech House, Boston, 1999)].

5

For the default gage length of 1 cm

6

Based on material properties of the standard sensor: polyimide-coated, low-bend-loss optical fiber. For temperatures up to 800°C,
contact Luna.