Luminex 100 IS Version 2.2 User Manual

Page 31

Advertising
background image

x

MAP

Technology

Acquisition Detail Tab

PN 89-00002-00-032 Rev. A

25

‹

To determine and set optimal gate settings:

1. Create a New Advanced batch with DD gate settings at 0 - 32767

(make sure all other settings are appropriate for your sample).

2. On plate layout choose a single well with the Acquire command

to obtain a single sample.

Note:

The sample is representative of the expected assay

samples. It consists of all bead sets combined, reagents, but no
patient samples. Gate positions are dependent upon buffer
composition. Any changes made to your assay buffer
composition may result in a different optimal gate location
.

3. Click and drag the right red line to a position on the right side of

the major peak to eliminate aggregate beads.

4. Click and drag the left red line to a position on the left side of the

major peak to eliminate small particles.

5. Note the numerical gate position indicated in the parenthesis in

the lower left corner of the histogram.

6. Use the numbered gate positions from step 5 in templates and

New Advanced Batches for assays with the same buffer
composition.

Note:

You may want to change the X axis of the Dot Plot for

troubleshooting purposes. We highly recommend using the
default settings in all other instances.

Table 6: DD Gate Relationships

Gate Width:

Wider

Narrower

%CV (within sample)

Higher

Lower

Sample Acquisition
Speed

Faster

Slower