Luminex LDS 1.7 SP1 to xPONENT 3.1 Rev 2 Upgrade Instructions User Manual

Page 14

Advertising
background image

Luminex LDS 1.7 SP1 Upgrade Instructions to xPONENT 3.1 - For In Vitro Diagnostic Use

10

LX100 Sample Probe

If you have a Luminex 200 (you can tell that it’s a Luminex 200 if
it has the sample probe shown in the next photo), click Upgrade
adjacent to the Detector Device text box. The Confirm Firm-
ware Download
dialog box opens. Click Yes to continue with the
upgrade.

LX200 Sample Probe

5. The Download In Progress dialog box opens displaying the

progress. Read the information presented on the dialog box.
When the download is finished the Status text box displays
Download Complete. Click OK.

6. Turn off the power to the analyzer.

7. Replace the serial-to-serial cable from the analyzer to the

computer with the new USB cable supplied in the upgrade kit by

Advertising