Addendum to User Manual Version 2.3

x

MAP Technology

2

PN 89-00002-00-140 Rev. A

Safety and

Regulatory Symbols

New safety and regulatory symbols have been added to the Luminex
100 system and manuals. Please become familiar with these symbols.

These symbols describe warnings, cautions, and general information
used in the operation of this instrument.

Calibrators and

Controls

Vortex the xMAP calibrators and controls containers to ensure
homogeneity. Do not dilute xMAP calibrator and control reagents.
Note that position of the vial may affect drop size. Be sure to invert
the vial such that it is perpendicular to the plate surface. When
calibrating the analyzer, load a microtiter plate with at least five
drops of each: CAL1 in well A1, CAL2 in well B1, CON1 in well
C1, CON2 in well D1 and distilled water in well E1 through H1 to
Wash a total of four times. You can use different wells as necessary.
To select different well locations in the software, click on the drop-
down arrow next to the entry cell for the calibrator or control, then
click in the well location on the microtiter plate image.

Symbol

Description

Symbol

Description

Symbol

Description

Alternating current
(ac)

Warning
(refer to manual)

Warning
(refer to manual)

Protective ground

Warning
(refer to manual)

Warning
(refer to manual)

On

Warning
(refer to manual)

Warning (refer to
manual)

Off

Catalog Number

Consult instructions
for use

In vitro Diagnostic
Medical Device

Batch Code

Serial Number

Temperature

Limitation

Expiration Date

Date of Manufacture

Manufacturer

Underwriter’s
Laboratory

Canadian
Underwriter’s
Laboratory

European Union
Conformity

WEEE - Do not
throw in trash

US