Meriam M201 Rotary Gas Meter Tester User Manual

Page 9

Advertising
background image

www.meriam.com

page 9 of 10

Addendum No. 9R124-IR

June, 2010

Instructions

1.

Install preferred connectors in the 1/8” NPT (female)

connections on the

M201 or M201 manifold. Use Teflon tape or

other suitable thread sealer.

2.

PTR and IPTR are normally open and the M201 sensor is

equalized.

The QTV valve must be manually moved to the

equalize position prior to

connecting pressure to the M201

.

3.

Make connections to the gas meter’s HI and LO ports. Be

certain to

observe the P1 (HI) and P2 (LO) on the M201.

4.

Turn M201 “On”. Display should read zero since HI and LO

sides are

connected and pressure is equalized. If not, re-zero

the M201 buy

simultaneously pressing the Min/Max and Hold

keys.

5.

To make differential pressure measurements with PTR and

IPTR

options, manually depress and hold the push-to-read

valve. To make

differential pressure measurements with QTR

option, manually move the

¼ turn valve to the closed position.

PTR Manifold

Valve is open and

DPVa

sensor

equalized until the p

ush button valve is

IPTR

Cartridge Valve

Integral valve is open

and DP sensor equalized

until the push button

valve is depressed

QTV Manifold

¼ Turn Valve is

manually opened to

equalize DP sensor and

closed to make readings

To protect the M201’s differential pressure sensor from
damage due to accidental over pressure, it is importand to
use the M201’s equalizing manifold correctly. Three
options are available: a Push-To-Read (PTR) valve
manifold, an Integral Push-To-Read (IPTR) cartridge valve
and a Quarter Turn Valve (QTV) manifold

Using the Equalizing Manifold Options