Table 1-1. specifications – Micromod Micro-DCI: 53IT5100A Indicator/Totalizer User Manual

Page 16

Advertising
background image

Table 1-1. Specifications

Item

Specification(s)

Analog Output (ANO0) Signal (is referenced
to power common)
Quantity

Signal Range

Load Range

Accuracy

1 (ANO0)

0 - 20 mA dc (4 - 20 mA dc typically)

0 - 750 ohms

±

0.2% of span

Switch Output (CCO0, CCO1) Signals (are
referenced to power common)
Quantity

Type

Configuration

Voltage

Current

2 (CCO0 and CCO1)

solid state switch output

solid state equivalent of a single pole single
throw, normally open or normally closed
contacts referenced to common.

30 V dc maximum

50 mA dc maximum

Datalink Communication

RS485, four wire, asynchronous; baud rates
between 300 and 28,800 inclusive.

Sampling and Update Attributes:
Program Scan Rate

Analog Input Signal Sampling Rate

Contact Input Signal Sampling Rate

Display Update

Output Signal Update

0.05 seconds

0.05 seconds

0.05 seconds

0.10 seconds

0.05 seconds

Environmental Characteristics
Controlled Environment

Ambient Temperature Limits

Relative Humidity Limits

Temperature Effects on Accuracy

Enclosed temperature controlled location (Class
A and B per ISA-S71.01 1985)

4 - 52

°

C (40 - 125

°

F)

10 - 90% maximum

±

0.28% per 28

°

C (50

°

F) change from reference

temperature 25

°

C (77

°

F)

Section 1. Introduction

1-11

Advertising