Uncertainty, Health physics division, Functional characteristics selectable units – Mirion Technologies SOR_RF User Manual

Page 2: Alarm levels up to one year battery life under, Normal conditions user selectable display modes, Backlit display, Exhaustive periodic self-testing, Training mode included, 1μgy to 10 gy gamma dose rate measurement range, Relative error of ambient measurement

Advertising
background image

Mirion Technologies (MGPI) Inc

5000 Highlands Parkway

Suite 150

Smyrna Georgia 30082

USA

T +1.770.432.2744

F +1.770.432.9179

Mirion Technologies (MGPI) SA

BP 1

F-13113 Lamanon

France

T +33 (0) 4 90 59 59 59

F +33 (0) 4 90 59 55 18

Mirion Technologies (RADOS) Oy

P.O. Box 506

FIN-20101 Turku

Finland

T +358 2 468 4600

F +358 2 468 4601

Mirion Technolgies (RADOS) GmbH

Ruhrstrasse 49

DE-22761 Hamburg

Germany

T +49 (0) 40 851 93-0

F +49 (0)40 851 93 256

Health Physics
Division

Since norms, specifi cations and designs are subject to occasional change, please ask for confi rmation of the information given in this publication.

SOR/RF

Active Dosimetry - Dosimeters

www.mirion.com

147349EN-A

FUNCTIONAL CHARACTERISTICS

Selectable units:

cGy; cGy/h; mSv; mSv/h; μSv; μS/h;

mrem and mrem/h

4 confi gurable dose and dose rate

alarm levels

Up to one year battery life under

normal conditions

User selectable display modes

Backlit display

Exhaustive periodic self-testing,

including the detector

Selectable Historical record of

measurements and events (3800

steps; 10 s; 1 min ;10 min; 1 h; 24 h)

Data storage in EEPROM

(qualifi ed > 10 years without battery)

Low battery (16 h left) pre-alarm and

alarm if the battery is removed

Neck lanyard or clip

Training mode included**

PHYSICAL CHARATERISTICS

Hp(10) dose equivalent measurements:

from 50 keV to 6 MeV

Ambient gamma dose measurement range:

1μGy to 10 Gy

Gamma dose rate measurement range:

from 0.1 μGy/h to 10 Gy/h

Gamma dose rate display:

from 1* or 10 μGy/h to 10 Gy/h

Saturation indication (above 10 Gy/h)

Relative error of ambient measurement:

<± 20 % over the dose measurement range

Energy response:

<± 20 % in the range 60 keV to 2 MeV

<± 50 % in the range 2 MeV to 6 MeV

Accredited factory calibration to IEC 17025

(COFRAC, procedure #127023)

Cs

137

(≈ 25 mSv/h), Am

241

(≈ 35 mSv/h),Co

60

(≈ 9 mSv/h)

conformity scope ±10 % Cs137 with

uncertainty

(K=2)

= ± 5 %

MECHANICAL CHARACTERISTICS

Dimensions: 80.4 x 48 x 9 mm (fl at housing)

(3.16 x 1.85 x 0.35 in)

Weight: 55 g (1.94 oz)

ENVIRONMENTAL CHARACTERISTICS

-20° to + 50°C (-4°F + 122°F) (normal operating range,

standard battery 3V LiMnO

2

CR2450)

Resistant to EMP and EMC (radars, GSM, jammers,...)

200 V/m pulse modulation, (

MiL-STD-461F RS103)

Resistant to water immersion (IP67: 1 m / 39.3 in),

drops, shocks,vibrations, low pressure,

Initial conditions, NBC environmental conditions

Complies with the following standards:

Meets MIL-STD-810 and MIL-STD-461 requirements

Qualifi ed by most of the NATO military laboratories

Complies with IEC 1283, ANSI 42-20 and NATO D104

NATO number: 6665 14 563 9423

NATO supplier n° F8929

Development status: Series production

* LLR : Low level Radiation

**a specifi c confi guration software is requi-

red for the XOM reader

SOR/RF inside the
arm-band pouch
(accessory)

SOR/RF version
to be clipped on a
pocket

SOR/RF worn round
the neck