Physical characteristics, Radiation detected: beta and gamma, Detector: dual large area silicon – Mirion Technologies NGM 204L User Manual

Page 2: Typical energy windows, Beta: 80 kev to 2.5 mev, Gamma: 80 kev to 2.5 mev, Typical measurement range, Kr: 3.7 10, Bq/m, Μci/cc)

Advertising
background image

Mirion Technologies (MGPI) SA

Route d’Eyguières
FR-13113 Lamanon
France

T +33 (0) 4 90 59 59 59
F +33 (0) 4 90 59 55 18

Mirion Technologies (MGPI) Inc

5000 Highlands Parkway
Suite 150
Smyrna, GA 30082
USA
T +1 770 432 2744
F +1 770 432 9179

Mirion Technologies (MGPI H&B) GmbH

Landsberger Strasse 328a
DE-80687 Munich
Germany

T +49 (0) 89515 13-0
F +49 (0) 89515 13 169

Mirion Commercial (Beijing) Co., Ltd.
Shanghai Jiangchang Commercial Branch

Room 801, 78 Jiangchang SanLu
Zhabei District, Shanghai 200436
PR of China
T +86 21 6180 6920
F +86 21 6180 6924

Since norms, specifi cations and designs are subject to occasional change, please ask for confi rmation of the information given in this publication.

Mirion, Mirion Technologies and MGP Instruments are trademarks and/or registered trademarks of Mirion Technologies, Inc. and/or its affi

liates in the United States and/or other countries.

All other trademarks are the property of their respective owners.

NGM

204L

Particulate, Gas & Iodine Monitors

PHYSICAL CHARACTERISTICS

Radiation detected: beta and gamma

Detector: dual large area silicon

Sampling chamber: 300 ml (300 cc)

Typical energy windows:

Beta: 80 keV to 2.5 MeV

Gamma: 80 keV to 2.5 MeV

Typical measurement range:

85

Kr: 3.7 10

+4

to 7.4 10

+13

Bq/m

3

(10

-6

to 2 10

+3

μCi/cc)

133

Xe: 3.7 10

+4

to 3.7 10

+12

Bq/m

3

(10

-6

to 10

+2

μCi/cc)

ENVIRONMENTAL CHARACTERISTICS

Long term temperature: +10°C to +40°C

(+50°F to +104°F)

Periodic temperature: -5°C to +55°C

(+23°F to +131°F)

MTBF: > 20 000 hours, with preventive maintenance

TID: 100 Gy (10

+4

rad)

PNEUMATIC CHARACTERISTICS

Standard fl ow rate: 35 l/min (1.24 scfm)

Pressure drop: according to the fi lter dust loading

MECHANICAL CHARACTERISTICS

Dimensions: 864 mm x 725 mm x 440 mm

(34 in x 28.5 in x 17.3 in)

Weight: ~ 220 kg (~ 495 lb)

Color: gray RAL 7030 (decontaminable paint)

Inlet tube connection: Ø 12 mm OD (1/2 in)

Outlet tube connection: Ø 12 mm OD (1/2 in)

ELECTRICAL CHARACTERISTICS

Power supply: 230 Vac – 50 Hz or 120 Vac – 60 Hz

Data link outputs: 1 RS232 and 2 isolated RS485

Alarm relays: 3 SPDT relays

I/O: 2 isolated analog outputs and 1 isolated analog

input (0/4-20 mA)

SIGNALING (Applicable to LPDU only)

Alphanumeric display: measurement, status…

Sound alarm: buzzer 90 dBA at 1 meter

Visual alarm: 3 lights (red, yellow, green)

REFERENCE STANDARDS

Nuclear: IEC60761

EMC: 2006/95/CE and 2004/108/CE, IEC61000-6-2

and IEC61000-6-4

VERSIONS

230 Vac or 120 Vac

LPDU or LPU

With or without dust fi lter holder

Pumping unit

Monitor wall or fl oor fi xing plate

ACCESSORIES

Calibration tools

Software

USB converters

www.mirion.com