Physical characteristics, Radiation detected: gamma, Detector: geiger muller tubes – Mirion Technologies GIM 210K User Manual

Page 2: Energy range: 80 kev to 2 mev, Typical measurement range, With two sg/gm 21 detectors: 10, Gy/h (10, Rad/h), With two sg/gm 22 detectors: 10, Rad/h) environmental characteristics

Advertising
background image

Since norms, specifi cations and designs are subject to occasional change, please ask for confi rmation of the information given in this publication.

Mirion Technologies (MGPI) SA

Route d’Eyguières

FR-13113 Lamanon

France

T +33 (0) 4 90 59 59 59

F +33 (0) 4 90 59 55 18

Mirion Technologies (MGPI) Inc

5000 Highlands Parkway

Suite 150

Smyrna, GA 30082

USA

T +1 770 432 2744

F +1 770 432 9179

Mirion Technologies (MGPI H&B) GmbH

Landsberger Strasse 328a

DE-80687 Munich

Germany

T +49 (0) 89515 13-0

F +49 (0) 89515 13 169

Mirion Commercial (Beijing) Co., Ltd.

Shanghai Jiangchang Commercial Branch

Room 801, 78 Jiangchang SanLu

Zhabei District, Shanghai 200436

PR of China

T +86 21 6180 6920

F +86 21 6180 6924

GIM 210K

Gamma & Neutronic Monitors

PHYSICAL CHARACTERISTICS

Radiation detected: gamma

Detector: Geiger Muller tubes

Energy range: 80 keV to 2 MeV

Typical measurement range:

With two SG/GM 21 detectors:
10

-6

to 10

-3

Gy/h (10

-4

to 10

-1

rad/h)

With two SG/GM 22 detectors:
10

-4

to 1 Gy/h (10

-2

to 10

+2

rad/h)

With one SG/GM 21 and one SG/GM 22 detector:
10

-6

to 1 Gy/h (10

-4

to 10

+2

rad/h)

ENVIRONMENTAL CHARACTERISTICS

Long term temperature:

+10°C to +40°C (+50°F to +104°F)

Periodic temperature (processing unit):
-5°C to +55°C (+23°F to +131°F)

Periodic temperature (detector):

-40°C to +70°C (-40°F to +158°F)

MTBF: > 20 000 hours

TID (processing unit): 100 Gy (10

+4

rad)

TID (SG/GM 21 detector): 6 Gy (6 10

+2

rad)

TID (SG/GM 22 detector): 370 Gy (3.7 10

+4

rad)

Protection index: IP65 and IK07

MECHANICAL CHARACTERISTICS

Dimensions:

Processing unit: 390 mm x 196 mm x 187 mm
(15.3 in x 7.7 in x 7.3 in)

Detector: 184 mm (7.2 in) x Ø 28 mm (1.1 in)

Weight:

Processing unit: 8.5 kg (18.7 lb)

Detector: 470 g (1 lb)

Color (processing unit): gray RAL 7030 (decontamina-
ble paint)

ELECTRICAL CHARACTERISTICS

Power supply: 230 Vac – 50 Hz or 120 Vac – 60 Hz

Data link outputs: 1 RS232 and 2 isolated RS485

Alarm relays: 3 SPDT relays

I/O: 2 isolated analog outputs and 1 isolated analog
input (0/4-20 mA)

SIGNALING (Applicable to LPDU only)

Alphanumeric display: measurement, status…

Sound alarm: buzzer 90 dBA at 1 meter

Visual alarm: 3 lights (red, yellow, green)

REFERENCE STANDARDS

Nuclear: IEC60532

Seismic: IEEE344 and IEC60980

EMC: 2006/95/CE and 2004/108/CE, EPRI 102323,
MIL STD 461 E, IEC61000-6-2 and IEC61000-6-4

VERSIONS

230 Vac or 120 Vac

LPDU or LPU

With or without RS485 junction box

Detector cable length:
from 10 m (32.8 ft) to 200 m (656 ft)

Junction box cable length:
2 m (6.56 ft), 5 m (16.4 ft) or 10 m (32.8 ft)

Detector calibrated in Gy/h or Sv/h

ACCESSORIES

Calibration tools

Software

USB converters

Seismic qualifi ed wall mounting bracket

www.mirion.com

144377EN-E

Advertising