MecHanIcal: caMera HeaD

Diameter

32mm (1.25”)

length (excluding cable taper)

340mm (13.4”)

Weight

450g (1 lb)

Weight with 38m (125’) cable

14kg (30 lb)

Housing

Polished 300 series
stainless steel

sPecIFIcatIons anD PerForMance:
electrIcal

camera tube

2

/

3

” Vidicon

Horizontal resolution

500 tv lines/raster height
Sensitivity for signal to noise
ratio of 20 db

scene Illumination
f/2.8 lens (2854°K)

40 lux (4fl) (f /2.8 SNR 20 db)

Grey scale

10 shades of grey, minimum

Video output

1V pp - 75 ohms

synchronization

EIA (525 lines, 60 fields/sec,
2:1 interlace) crystal controlled,
internal option to CCIR standard
(625 lines, 50 fields/sec,
2:1 interlace)

sPecIFIcatIons anD PerForMance: PoWer

line Voltage

90–110V; 100V–120V; 200–240V

line Frequency

60/50 Hz ± 5%

camera

55 VA nominal

total with accessories

Up to 325 VA

sPecIFIcatIons anD PerForMance: oPtIcal

lens

16mm f/2.8 radiation tolerant

Diagonal Field of View

38° in air, 28° in water

remote focus range

straight Viewing
(no attachments)

25mm (1”) to infinity

right angle attachments

From contact to infinity

oPtIons

• Detachable camera head and cable assembly
• Rack mount CCU

MecHanIcal: caMera control unIt

Width

478mm (18.8”)

length

305mm (12”)

Height

157mm (6.2”)

Weight

15kg (33 lb)

Video connectors

BNC

MecHanIcal: stanDarD caBle

length

38m (125’)

Diameter

14.5mm (0.57”)

Minimum Bend radius

150mm (6”)

enVIronMental: caMera HeaD

operating temperature

-25°C TO 60°C (-13°F to 140°F)

storage temperature

-30°C to 71°C (-22°F to 160°F)

operating Depth

Up to full cable length (38m; 125’)

cumulative Dose

1MGy (H

2

0) [

60

Co] (10

8

rads)

Vidicon

chalnicon/
newvicon

Gamma Dose rate

20kGy/hour
2 x 10

6

rads/hour

2kGy/hour
2 x 10

5

rads/hour

enVIronMental: caMera control unIt

operating temperature

-18°C to 50°C (0°F to 122°F)

storage temperature

-30°C to 71°C (-22°F to 160°F)

relative Humidity

0 to 95% (non condensing)