Operating  Modes  

The nanoZ application has five separate modes of operation for performing a
variety of different tasks. These modes can be selected by clicking a button on
the left pane of nanoZ application or via the Mode menu on the main menu.

The following instructions assume that a suitable electrode adaptor and probe
configuration have been chosen, and the electrode sites of interest have been
selected in the Probe window.

The screenshots below were taken from Windows 7. The appearance of the user
interface may vary depending on the version of Windows you are using.

Manual  control  mode  

This mode provides manual control of the nanoZʼs channel selection, impedance
testing, and DC constant current electroplating functions.


Selecting Impedance (A) will continuously measure the impedance of the
currently selected channel (B) at the specified Test freq (C). The impedance (in
Mohms), and phase angle (in degrees) will be displayed on the Meter. If the test
signal is saturated (e.g. high levels of 50 or 60Hz noise) the ʻTest signal clippedʼ
warning message will be displayed and the Meter will show the last valid
impedance measurement.

If One shot is checked, a single impedance measurement will be made. To stop
continuous impedance measurements, select Off (A).