Ecoflexmea36 – Multichannel Systems EcoFlexMEA36 User Manual

Page 2

Advertising
background image

Multi Channel Systems
MCS GmbH
Aspenhaustrasse 21
72770 Reutlingen
Germany

Fon +49-7121-9 09 25-

0

Fax +49-7121-9 09 25-11

[email protected]
www.multichannelsystems.com

© 2013 Multi Channel Systems MCS GmbH

Product information is subject to change without
notice.

EcoFlexMEA36

Electrode layout

GND 2

GND 1

29

31

1

3

27

28

30

2

4

5

25

26

32

8

6

7

23

22

24

16

10

9

21

20

18

14

12

11

19

17

15

13

REF

REF

50

m

GND 1 is a large ground electrode connected to pin 1 of the MPA32I input connector. GND 2 is a second
ground electrode connected to pin 36. The REF electrodes are reference electrodes connected to pin 2
and 35, respectively. Both ground inputs and both reference electrode inputs are equal, that is, they are
connected to each other inside the standard MPA32I. Please see the MPA32I manual for details.

The numbers in the electrodes are the recording channel numbers that refer to the channel numbers
of the data acquisition program. Please make sure that you have selected “Configuration” in the “Channel
Layout” under “Data Source Setup” with a total number of 64 channels. In “Amplifier”, please choose
FA32I/S or FA64I/S and in “MEA” EcoFlexMEA36. In Layout tab of the display, please click ”Default Map”.

The letter digit code below is the electrode identifier and refers to the position of the electrode in the grid.

Important: The channel map is constructed by looking on the back side of the electrodes,
because the FlexMEA electrodes are placed on the preparation upside down!

Warning: The device may only be used together with the MPA32I from Multi Channel Systems
MCS GmbH, and only for the specified purpose. Damage of the device and even injuries can
result from improper use.

!

300

m

A2 A3 A4 A5

B1 B2 B3 B4 B5 B6

C1 C2 C3 C4 C5 C6

D1 D2 D3 D4 D5 D6

E1 E2 E3 E4 E5 E6

F2 F3 F4 F5

300

m

If you use a MEA64-System with more than one MPA32I, the signal collector SC2x32 leads the output
channels of the second amplifier to channel numbers 33 to 64. Please read the SC2x32 data sheet.

Advertising