Multichannel Systems EcoFlexMEA24 User Manual

Ecoflexmea24

Advertising
background image

Multi Channel Systems
MCS GmbH
Aspenhaustrasse 21
72770 Reutlingen
Germany

Fon +49-7121-9 09 25- 0
Fax +49-7121-9 09 25-11

[email protected]
www.multichannelsystems.com

© 2013 Multi Channel Systems MCS GmbH

Product information is subject to change
without notice.

EcoFlexMEA24

Flexible Microelectrode Array with 24 electrodes for use with 32-Channel Miniature Preamplifer
MPA32I for in vivo or in vitro applications.

Technical Specifications EcoFlexMEA24

Temperature compatibility

10 - 125 °C

Dimension (W x D)

36 mm x 25 mm

Thickness (region of electrodes)

50 μm

Base

material

Polyimide

(Kapton)

Contact

pads

and

leads

Gold

Electrode

diameter

80

μm

Interelectrode distance (centre to centre)

300 μm

Electrode

heights Planar

Electrode

type

Gold

Isolation

type

Polyimide

(Kapton)

Electrode impedance

Approximately 30 - 50 k

Electrode layout grid

10 x 2 + 4

Number of recording electrodes

23

Number of reference electrodes

2 internal reference electrodes

Number of ground electrodes

2 ground electrodes

MC_Rack

Version 4.1.1 and higher

Data Source Setup

Source Layout: Configuration, Total Number of Channels: 64,

Amplifier:

FA32I/S

or

FA64I/S,

MEA:

EcoFlexMEA24.

Channel map

Display: Layout tab > Default Map (EcoFlexMEA24.cmp).

Cleaning

Rinse with distilled water. EcoFlexMEAs are heat stable and can be autoclaved.

1......35 Pin Top GND

2.....36 Pin Bottom Electrodes

Top
35 33 31 29 27 25 23 21 19 17 15 13 11 9 7 5 3 1

36 34 32 30 28 26 24 22 20 18 16 14 12 10 8 6 4 2
Bottom

Top

Connect the EcoFlexMEA24 directly to a 32-Channel Miniature Preamplifier.
The electrode field is at the bottom of the EcoFlexMEA24.
Insert the EcoFlexMEA24 into the MPA32I with the electrode field up.

Advertising