Multichannel Systems 60MEA500_10iR-Ti User Manual

Multichannel Systems Equipment

Advertising
background image

Multi Channel Systems
MCS GmbH
Aspenhaustrasse 21
72770 Reutlingen
Germany

Fon +49-7121-9 09 25- 0
Fax +49-7121-9 09 25-11

[email protected]
www.multichannelsystems.com

© 2013 Multi Channel Systems MCS GmbH

Product information is subject to change
without notice.

60MEA500/10iR-Ti

60MEA500/10iR-Ti, 60MEA500/30iR-Ti

Technical Specifications Micro Electrode Array with 6 x 10 Layout

Temperature compatibility

0 - 125 °C

Dimension (W x D x H)

49 mm x 49 mm x 1 mm

Base

material

Glass

Track

material

Ti

(Titanium)

Contact

pads

TiN

(Titanium

nitride)

Electrode

diameter

10

or

30

μm

Interelectrode distance (centre to centre)

500 μm

Electrode

height Planar

Electrode

type

TiN

(Titanium

nitride)

Isolation

type

Silicon

nitride

500

nm

(PEVCD)

Electrode

impedance

30

-

50

k

for 30m electrodes,

250

-

400 k

for 10m electrodes

Electrode

layout

grid

6

x

10

Number of recording electrodes

59

Number of reference electrodes

1 internal reference electrode (iR)

MC_Rack
Source layout in “Data Source Setup”

2 dim. (MEA)

Channel

map

MEA-500_6x10.cmp

MEA perfusion chamber

(w/o) Without ring

(gr)

Glass ring: ID +/- 19 mm, OD 24 mm, height 6 / 12 mm

(pr)

Plastic ring without thread: ID 26.5 mm, OD 30 mm, height 6 / 3 mm

(pr-T) Plastic ring with thread: ID 26 mm, OD 30 mm, height 6 / 15 mm

MEA with electrode layout grid 6 x 10

Advertising