Technical Specifications 60HexaMEA

Temperature compatibility

0 - 125 °C

Dimension (W x D x H)

49 mm x 49 mm x 1 mm

Thickness (region of electrodes)

Base

material

Glass

Contact pads

Indium tin oxide (ITO) or Titanium nitride (TiN)

Track material

Indium tin oxide (ITO) or Titanium (Ti)

Electrode

diameter

10,

20,

30

μm

Interelectrode distance (centre to centre)

30, 60, 90 μm

Electrode

height Planar

Electrode

type

Titanium

nitride

(TiN)

Isolation

type

Silicon

nitride

500

nm

(PEVCD)

Electrode

impedance

30

k

 - 50 kfor 30 μm electrodes,

250

-

400

k

 for 10 and 20 μm electrodes

Electrode

layout

grid

hexagonal

Number of recording electrodes

60

Number of reference electrodes

without internal reference electrode

MC_Rack:
Source Layout in Data Source Setup

2 dim. (MEA) or Configuration

MCS

Channel

map

HexaMEA.cmp

MEA perfusion chamber

(w/o) Without ring

(gr)

Glass ring: ID +/- 19 mm, OD 24 mm, height 6 / 12 mm

(pr)

Plastic ring without thread: ID 26.5 mm, OD 30 mm, height 6 / 3 mm

(pr-T) Plastic ring with thread: ID 26 mm, OD 30 mm, height 6 / 15 mm

60 Hexa Microelectrode Array

30 60 90 90 μm

10 20 30 μm