Multichannel Systems MEA1060 Amplifiers User Manual

Mea1060 amplifiers

Advertising
background image

Technical Specifications

Date of Print: 26.01.2007

MEA1060-Inv-Standard

MEA1060-Up-Standard

MEA1060-Inv-BC-Standard (incl. FA60S-BC) MEA1060-Up-BC-Standard (incl. FA60S-BC)

MEA96-Well-Standard

Operating temperature

10 °C to 50 °C

10 °C to 50 °C

10 °C to 50 °C

10 °C to 50 °C

10 °C to 50 °C

Storage temperature

0 °C to 50 °C

0 °C to 50 °C

0 °C to 50 °C

0 °C to 50 °C

0 °C to 50 °C

Relative humidity

10 % to 85 %

Dimensions (W x D x H)

165 mm x 165 mm x 20 mm (29 mm with
sockets)

165 mm x 165 mm x 14.5 mm (19 mm with
sockets)

165 mm x 165 mm x 20 mm (29 mm with
sockets)

196 mm x 184 mm x 17.5 mm

Weight

800 g

600 g

750 g (without FA60S) + 450 g (FA60S)

750 g (without FA60S) + 450 g (FA60S)

Supply voltage

± 6 VDC to ± 9 VDC

± 6 VDC to ± 9 VDC

± 6.5 VDC to ± 9 VDC

± 6.5 VDC to ± 9 VDC

± 6.5 VDC to ± 9 VDC

Supply current

max. 220 mA, typically ± 150 mA

max. 220 mA, typically ± 150 mA

± 550 mA

± 550 mA

max. 220 mA, typically ± 150 mA

Number of stimulus input channels

60

60

2

2

0

Maximum stimulus input voltage

n/a

n/a

± 5 V

± 5 V

n/a

Number of input channels

60

60

60

60

96

Input voltage

± 4 mV (with respect to the standard gain of
1200)

± 4 mV (with respect to the standard gain of
1200)

± 4 mV (with respect to the standard gain of
1100)

± 4 mV (with respect to the standard gain of
1100)

± 4 mV (with respect to the standard gain of
1000)

Input impedance

10

11

Ω parallel to 10 pF

10

11

Ω parallel to 10 pF

10

11

Ω parallel to 10 pF

10

11

Ω parallel to 10 pF

10

11

Ω parallel to 10 pF

Input noise

< 800 nV

RMS

< 800 nV

RMS

< 800 nV

RMS

< 800 nV

RMS

< 800 nV

RMS

Noise density

e

n

= 15

e

n

= 15

e

n

= 15

e

n

= 15

e

n

= 15

Number of output channels

60

60

60

60

96

Output voltage

± 5 V

± 5 V

± 5 V

± 5 V

± 5 V

Output current

10 mA

10 mA

10 mA

10 mA

10 mA

Output impedance

300

300

300

300

300

Bandwidth*

1 Hz to 3 kHz (standard), 1 Hz to 10 kHz
(custom)

1 Hz to 3 kHz (standard), 1 Hz to 10 kHz
(custom)

1 Hz to 3 kHz (standard), 1 Hz to 8 kHz
(custom)

1 Hz to 3 kHz (standard), 1 Hz to 8 kHz
(custom)

0.6 Hz to 2.2 kHz

Filter slope

60 db/decade

60 db/decade

MEA1060-BC-PA: 20 db/decade (high pass), 60
db/decade (low pass); FA60S-BC: 20 db/decade
(high pass), 80 db/decade (low pass)

MEA1060-BC-PA: 20 db/decade (high pass), 60
db/decade (low pass); FA60S-BC: 20 db/decade
(high pass), 80 db/decade (low pass)

60 db/decade (low pass), 40 db/decade (high
pass)

Gain

1200 (standard), 500 to 5000 (custom)

1200 (standard), 500 to 5000 (custom)

1100 (standard), 500 to 5000 (custom)

1100 (standard), 500 to 5000 (custom)

1000

Typical switch enable time

n/a

n/a

20 ns

20 ns

n/a

Heating element (HE)

(separate lid and bottom heating)

Heating temperature

Ambient temperature (min 10 °C) to 50 °C

Ambient temperature (min 10 °C) to 50 °C

Ambient temperature (min 10 °C) to 50 °C

Ambient temperature (min 10 °C) to 50 °C

Ambient temperature (min 10 °C) to 50 °C

Accuracy

0.1 °C

0.1 °C

0.1 °C

0.1 °C

0.1 °C

Recovery time

0.5 min to 2 min

0.5 min to 2 min

0.5 min to 2 min

0.5 min to 2 min

0.5 min to 2 min

Typical time constant

50 s

50 s

50 s

50 s

50 s

Standard hole size

8 mm (HE-Inv-8)

17.6 mm (HE-UP)

8 mm (HE-Inv-8)

17.6 mm (HE-UP-BC)

n/a

Optional hole sizes

12 mm (HE-Inv-12), 23 mm (HE-Inv-23)

-

12 mm (HE-Inv-12), 23 mm (HE-Inv-23)

-

n/a

Input voltage

max 10 V

max 10 V

max 10 V

max 10 V

max 10 V

Input current

max 2 A

max 2 A

max 2 A

max 2 A

max 2 A

Temperature sensor

PT 100

PT 100

PT 100

PT 100

PT 100

Resistance

15 ± 2

15 ± 2

15 ± 2

15 ± 2

15 ± 2

Thermal resistance

6 °C / Watt

6 °C / Watt

6 °C / Watt

6 °C / Watt

6 °C / Watt

Calibration constant

T = c * (R-R

0

)/R

0

for c = 259.7 °C

T = c * (R-R

0

)/R

0

for c = 259.7 °C

T = c * (R-R

0

)/R

0

for c = 259.7 °C

T = c * (R-R

0

)/R

0

for c = 259.7 °C

T = c * (R-R

0

)/R

0

for c = 259.7 °C

Microscope compatibility

Inverse microscopes

Upright** microscopes

Inverse microscopes

Upright** microscopes

n/a

Distance focal plane to microscope table

3.5 mm

8 mm

3.5 mm

9.5 mm

MEA1060 Amplifiers

MEA1060 Amplifiers can be ordered with any gain and bandwidth configurations by the user's choice. The name specifies the configuration. The first number after the microscope type specifies the gain, followed by the lowest frequency, and followed by the highest frequency.
For example, MEA1060-Inv-5000-300-3000 stands for a MEA1060 amplifier for inverted microscopes with a gain of 5000, and 300 Hz to 3 kHz bandwidth. MEA1060-BC operates software-controlled and features a blanking circuit that prevents stimulus artifacts.

MEA1060 amplifiers are available for inverted microscopes and upright microscopes, MEA1060-INV and MEA1060-UP, respectively. **= MEA1060-UP amplifiers were optimized for upright microscopes, but can also be used for inverted microscopes, provided that the focal plane is
compatible with the microscope objective. This is useful whenever an excellent accessibility of the MEA from the top is required. With an MEA1060-UP, the MEA culture chamber is the only limiting parameter for accessibility. The opening in the base plate of the amplifier is circular and
hemispherical to follow the dimensions of an objective. The largest opening at the bottom is 35 mm in diameter.

MEA1060 Amplifiers are for use with two-dimensional 60-electrode microelectrode arrays (MEA). MEA 96-Well Amplifier is for use with 96-well plates with integrated recording electrodes (QT-Well plates).

Hz

nV /

Hz

nV /

Hz

nV /

Hz

nV /

Hz

nV /

Page 1 of 1