Multichannel Systems PH01 User Manual

Ph01, Perfusion cannula for controlled fluid temperature

Advertising
background image

Multi Channel Systems
MCS GmbH
Aspenhaustrasse 21
72770 Reutlingen
Germany

Fon +49-7121-9 09 25- 0
Fax +49-7121-9 09 25-11

[email protected]
www.multichannelsystems.com

2013 Multi Channel Systems MCS GmbH

Product information is subject to change
without notice.

Scope of delivery:
1 PH01
1 Cable (1.8 m) with connectors ready to be used with TC01 /TC02 controller.

PH01

Perfusion Cannula for Controlled Fluid Temperature

Operating temperature

10 °C to 40 °C

Storage temperature

0 °C to 50 °C

Dimensions (W x D x H)

111 mm x 10.5 mm x 24 mm (incl. cannula)

48.5 mm x 10.5 mm x 24 mm (w\o cannula)

Weight 17

g

Cannula

type stainless

steel

outside

diameter

1

mm

inside

diameter

0.7

mm

Integrated sensor:

Type

Pt 100 1/3 DIN B

Accuracy

+/-

0.1

°C

Measuring Method

four wire measuring bridge

Heating Element:

Resistance

22

Ohm

Heating voltage

max. 12 Volt

Heating temperature

ambient temperature to 50 °C

Setting time

0.5 min to 5 min