Multichannel Systems pMEA32S12 Layout 2 User Manual

Multichannel Systems Equipment

Advertising
background image

Multi Channel Systems
MCS GmbH
Aspenhaustrasse 21
72770 Reutlingen
Germany

Fon +49-7121-9 09 25- 0
Fax +49-7121-9 09 25-11

[email protected]
www.multichannelsystems.com

© 2013 Multi Channel Systems MCS GmbH

Product information is subject to change
without notice.

pMEA32S12 Layout 2

Technical Specifications pMEA32S12 Layout 2

Temperature compatibility

10 - 50 °C

Dimension (W x D x H)

49 mm x 25 mm x 1.8 mm

Base material

Polyimide foil on ceramic carrier with perforation

Perforation:
Total area of holes

0.8 mm

Diameter of the holes

90, 75, 50, 30, 20 μm

Track

material

Ti

(Titanium)

Contact

pads

TiN

(Titanium

nitride)

Electrode

diameter

30

μm (recording electrodes) 50 μm (stimulation electrodes)

Interelectrode distance (centre to centre)

90 μm and 150 μm (recording electrodes),

90 μm and 150 μm (stimulation electrodes)

Electrode

height Planar

Electrode

type

TiN

(Titanium

nitride)

electrodes

Isolation

type

Polyimide

foil

Electrode impedance

Approximately 30 - 50 k

Electrode layout grid

1 x 10 + 1 x 12 + 1 x 10 (recording electrodes),

6

x

2

(stimulation

electrodes)

Number of recording electrodes

32

Number of stimulation electrodes

12

Number of reference electrodes

1 internal reference electrode

MC_Rack “Source layout”

Configuration (MEA2100-32)

in “Data Source Setup”

1 dimensional, no digital channel (USB-MEA32-STIM4)

Channel

map

pMEA-32S12-L2_12x3.cmp

Cleaning Rinse

with

distilled

water.

Do

not

use

ultrasonic

bath!

Do not autoclave or sterilize pMEAs by heat.

These MEA types are not heat-stable and will be rreversibly damaged!

pMEA perfusion chamber

(w/o) Without ring

(gr) Glass ring: ID +/- 19 mm, OD 24 mm, height 6 / 12 mm

2

Perforated MEA for use with MEA2100-32- and USB-MEA32-STIM4-System

Advertising