Multichannel Systems Test-256MEA User Manual

Test-256mea, Test model probe for usb-mea256-systems

Advertising
background image

Multi Channel Systems
MCS GmbH
Aspenhaustrasse 21
72770 Reutlingen
Germany

Fon +49-7121-9 09 25- 0
Fax +49-7121-9 09 25-11

[email protected]
www.multichannelsystems.com

© 2011 Multi Channel Systems MCS GmbH

Product information is subject to change
without notice.

Scope of delivery:
1 Test Model Probe Test-256MEA
1 CB-GND cable (CB-GND cable with a length of ca. 50 mm and 0.76 mm connectors at each end).

Test-256MEA

Warning: The Test-256MEA may only be used together with the USB-MEA256-System
from Multi Channel Systems MCS GmbH, and only for the specified purpose. Damage
of the device and even fatal injuries can result from improper use.

!

For the intention to check the noise level of the electrodes of an USB-MEA256-System without
external signals, please connect “signal source” and “signal ground” sockets with a CB-GND cable
for a short circuit. All 252 recording electrodes are internally connected to the resitor-capacitor pair.

To feed in test signals into the USB-MEA256-System via the test model probe, please connect
the signal source, for example a STG, to the “signal source” socket, and the signal ground
via “signal ground” socket to the amplifiers ground.

For ground connection and for short circuit please use the provided cable CB-GND.

The impedance between signal source and contact pads averages 120 k

 at 1 kHz.

Test Model Probe for USB-MEA256-Systems

Signal source

Signal ground, internal connected
to the four GND points.

GND

GND

GND

GND

The provided test model probe simulates a 256MEA with a resistor of 220 k

 and a 1 nF capacitor

between ground and electrode. It can be used for testing the noise level of the electrodes of an
USB-MEA256-Systems for all 252 recording electrodes simultaneously.

GND = “bath” Pad = “electrode”

220 k

1 nF