48

Chapter 2

Basic Operation
Configuring the RF Output

editing several points in the List Mode Values table. For information on using tables, see

“Using Table

Editors” on page 38

.

1.

Press

Sweep Repeat Single Cont

.

This toggles the sweep repeat from continuous to single. The SWEEP annunciator is turned off. The
sweep will not occur until it is triggered.

2.

Press

Sweep Type List Step

.

This toggles the sweep type from step to list.

3.

Press

Configure List Sweep

.

This opens another menu displaying softkeys that you will use to create the sweep points. The
display shows the current list data. (When no list has been previously created, the default list
contains one point set to the signal generator’s maximum frequency, minimum amplitude, and a
dwell time of 2 ms.)

4.

Press

More (1 of 2)

>

Load List From Step Sweep

>

Confirm Load From Step Data

.

The points you defined in the step sweep are automatically loaded into the list.

To Edit List Sweep Points

1.

Press

Return

>

Sweep

>

Off

.

Turning the sweep off allows you to edit the list sweep points without generating errors. If sweep
remains on during editing, errors occur whenever one or two point parameters (frequency, power,
and dwell) are undefined.

2.

Press

Configure List Sweep

.

This returns you to the sweep list table.

3.

Use the arrow keys to highlight the dwell time in row 1.

4.

Press

Edit Item

.

The dwell time for point 1 becomes the active function.

5.

Press

100

>

msec

.

This enters 100 ms as the new dwell time value for row 1. Note that the next item in the table
(in this case, the frequency value for point 2) becomes highlighted after you press the terminator
softkey.

6.

Using the arrow keys, highlight the frequency value in row 4.

7.

Press

Edit Item

>

545

>

MHz

.

This changes the frequency value in row 4 to 545 MHz.

8.

Highlight any column in the point 7 row and press

Insert Row

.

This adds a new point between points 7 and 8. A copy of the point 7 row is placed between
points 7 and 8, creating a new point 8, and renumbering the successive points.

9.

Highlight the frequency item for point 8, then press

Insert Item

.

Pressing

Insert Item

shifts frequency values down one row, beginning at point 8. Note that the

original frequency values for both points 8 and 9 shift down one row, creating an entry for point
10 that contains only a frequency value (the power and dwell time items do not shift down).